background image

26

27

Spindle: 
In this menu, you can select a required spindle for measuring�
1� List: According to the rough viscosity of the media to select a required spindle from the spindle list under 

different spindle groups (Standard spindle, Spiral, ELVAS���)� A tick (√) show the selected spindle and group�

2� Search: According to required speed and rough viscosity of the media to select a required spindle�

Note:

 The maximum viscosity (full scale range) can be automatically calculated basing on the selected spin-

dle, the speed and the integrated spring series (lo-vi/me-vi/hi-vi I/hi-vi II)�

   

Temperature:

Probe Temperature: 
In this menu, you can check the temperature of the media when the temperature sensor is connected�

Display: 
In this menu, you can specify the measured temperature value on the screen� A tick (√) show that the option 
is activated�

Calibration: 
In this menu, you can calibrate the connected temperature sensor with help of an external standard reference 
temperature thermometer� 

Note:

 The temperature sensor must be calibrated whenever it is replaced or changed, or if a new temperature 

sensor is installed� The temperature sensor supplied with the system is calibrated in factory�

   

Display:

Temperature: 
In this menu, you can specify the measured temperature value on the screen� A tick (√) show that the option 
is activated�

Level meter: 
In this menu, you can check the levelling status of the device�

  

Graph:

Axis Assignment: 
In this menu you can select 2 parameters from the 7 options (shear rate, shear stress, speed���) for axis 
assignment� The ticks (√) show that the options that are activated� The selected parameters will be used for 
the graph in the profile�

Axis scale:
When parameter “Time“ is not selected, the menu option “Axis scale“ is disabled and shown in grey color�
When you select “Time“ as one axis, the “Axis scale” gets enabled and can be edited� You can select 15, 30, 
60 or 120 minutes as the axis scale� The selected scale will be used for the graph in the profile�

   

Program:

Program: 
In this menu, 5 user-defined “speed-time” profiles can be created� A program can consist of up to 10 segments� 
Once a program has been selected, the following options are available:

1� Start:

Starts the program upon request of loop mode�

Infinite loop:

 

Upon completion of the last segment, the program continues with the first segment until 

the user ends the program by stopping a device function�

Loop count:

 

Indicates the total number of loop cycles until program end�

When the selected program is started by pressing on menu option “Start” with rotary/push knob, screen 
reminds the user to confirm the program (PR) control� Press “OK” with rotary/push knob to start the 
program control�

Note: 

At the end of the program all device functions are switched off�

2� Edit:

Edit/change the selected program by pressing on the button “Edit“ with rotary/push knob�
Start to edit the selected program parameters (speed and time ) by pressing on the selected program� You 
can “Edit”, “Delete“, “Insert“ and “Save“ the program segment with the rotary/push knob�
When you edit the program for at least one segment, symbol (   ) for respective program will appear�

In a program, user can define up to 10 program segments� The selected program segment is highlighted 
in blue color� Then, the user can edit, delete, insert or save segment in this program by rotating/pressing 
on the rotary/push knob� 
When editing a parameter (Speed / Time), press the rotary/push knob on the selected parameter� The 
highlighted cursor changed into yellow color indicate that parameter can be edited by rotating the rotary/
push knob�
When the “Back” button is pressed after editing, the program is saved after pressing on “OK“ as reminding 
in the screen�

3� Delete: 

Delete the selected program by pressing on the button “Delete“ with rotary/push knob�
If a selected program is deleted by pressing on menu option “Delete” with rotary/push knob, all the program 
parameters will be emptied� The symbol (   ) disappears�

4� View: 

View the “Speed (rpm)-Timer (hh:mm:ss)“ graph of the selected program by pressing on the button “View“ 
with rotary/push knob�

5� Rename: 

Rename the selected program by pressing on the button “Rename“ with rotary/push knob�

Summary of Contents for ROTAVISC

Page 1: ...IKA ROTAVISC...

Page 2: ...ENGLISH...

Page 3: ...on and leveling 17 Device self checking 18 Place the beaker and tempering bath in proper position 19 Select spindle 19 Assemble the selected spindle 20 Set the test speed 0 200 rpm 22 Start the test...

Page 4: ...ained staff work with the device Follow the safety instructions guidelines occupational health safety and accident prevention regulations The device must only be used in a technically perfect conditio...

Page 5: ...isposal instructions The device packaging accessories and batteries must be disposed of in accordance with national regulations Use The IKA ROTAVISC must be used in conjunction with the IKA ROTASTAND...

Page 6: ...O P Q R S M ROTASTAND pillar N ROTASTAND base O Screw washer P Hexagon socket screw M8 Q Hexagon socket screw M4 R Screwdriver AF 2 5 mm DIN ISO 2936 S Screwdriver AF 5 mm DIN ISO 2936 Assembly 1 2 1...

Page 7: ...1 2 1 2 Click Push 8 9 24V 100 mA LPS 1 2 3 Note The protection cap protects the coupling nut and the inside pivot from damage during transportation Keep the protection cap in the hole of the stand in...

Page 8: ...button Return to previous screen Menu button Touch it once Enter main menu screen Touch it a second time Back to working screen Control knob Rotating Pressing Start Stop the motor for measuring press...

Page 9: ...symbols on the working screen Symbol Designation Function Profile symbol Indicate the selected profile Accurate mode symbol Indicate the accurate mode is selected Balanced mode symbol Indicate the ba...

Page 10: ...pa s Accuracy 22 mpa s Spindle SP 4 Speed 28 20 rpm Maximum Viscosity 226950 mpa s Accuracy 2269 mpa s ON O O ON N N N O ON O ON O O ON N N N O ON O ON O O ON N N N O ON O 2 ON O O ON N N N O ON O 1 2...

Page 11: ...ity Spindle Search Spindle SP 1 Speed 28 20 rpm Maximum Viscosity 2 269 mpa s Accuracy 22 mpa s None Profile 1 SP 1 Stop 0 mPa s rpm n 0 00 00 00 00 0 0 M y Accuracy y Viscosity Spindle Search Spindle...

Page 12: ...dit the values or settings or activate deactivate a function Rotate the control knob to OK or press the Back button or Menu button to end the procedure and return to the previous menu Note The menu op...

Page 13: ...Visc Display Firmware Update Screen Background Black White 100 Sound Volume Factory Settings Communication Device Name ROTAVISC Information Viscosity Software Control mode Density Version Spindle 1 0...

Page 14: ...ram Program In this menu 5 user defined speed time profiles can be created A program can consist of up to 10 segments Once a program has been selected the following options are available 1 Start Start...

Page 15: ...e the control mode as Accurate Balance or Fast For details see Accu racy under menu option Viscosity 2 Start The user can define the start speed 0 01 to 200 rpm for spindle 3 End The user can define t...

Page 16: ...ika com ika lws download usb driver zip Install the driver by running the setup file Then connect the IKA device through the USB data cable to the PC The data communication is via a virtual COM port R...

Page 17: ...the device for repair including a short description of the fault Maintenance and cleaning The device is maintenance free It is only subject to the natural wear and tear of components and their statis...

Page 18: ...ion Connection for external temperature sensor PT 100 Working temperature C 100 300 Working temperature display TFT LCD Temperature measurement resolution K 0 1 Temperature measurement accuracy 100 C...

Page 19: ...s Spindle lo vi me vi hi vi l hi vi II SP 1 6 000 64 000 128 000 512 000 SP 2 30 000 320 000 640 000 2 560 000 SP 3 120 000 1 280 000 2 560 000 10 240 000 SP 4 600 000 6 400 000 12 800 000 51 200 000...

Page 20: ...Materials Determination of Viscosity of Crude or Modified Isocyanates D 5018 89 Standard Test Method for Shear Viscosity of Coal Tar and Petroleum Pitches Reapproved 1999 D 5133 01 Standard Test Meth...

Page 21: ...00 eMail info ika in UNITED KINGDOM IKA England LTD Phone 44 1865 986 162 eMail sales england ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guangz...