22

 Setting the temperature control mode

When using an external PT 1000 temperature sensor, the user can 
choose between two types of control:

PI mode
Good control results, minimized overshooting, slow rise in temperature

2P-mode (two-point controller)
Maximum heating rate, increased overshooting

Factory setting: PI mode

Changing the temperature control mode
1. Set target temperature to 2 °C
2. Switch off device
3. Hold down temp knob
4. Switch on device
This activates or deactivates the two-point controller (2P) for the 
external PT 1000 temperature sensor (2P appears on the display).
When switching on the device, 2P is also displayed next to the 
operating mode A/B/D when the two-point controller is activated.
When the heating function is active in 2P mode, (with external PT 1000 
temperature sensor) the actual temperature flashes - that notifies the 
user that the process temperature has been greatly overshot!

 Maintenance

The device is maintenance-free.

Cleaning

For cleaning disconnect the main plug.

Only use cleansing agents which have been recommended by 

IKA

®

.

Use to remove:
Dyes isopropyl 

alcohol

Construction materials  isopropyl alcohol/water containing surfactant

Cosmetics 

isopropyl alcohol/water containing surfactant

Foodstuffs 

water containing surfactant

Fuels 

water containing surfactant

-  Do not allow moisture to get into the appliance when cleaning.
-  Wear protective gloves when cleaning the devices.
-  Please consult with 

IKA

®

 before using any cleaning or deconta-

  mination methods, other than those recommended here.

Ordering spare parts
When ordering spare parts, please give:
- Machine type
- Manufacturing number, see type plate
- Item number and designation of the spare part,
  see 

www.ika.com

.

Repair

Please send instrument in for repair only after it has been cleaned 
and is free from any materials which may constitute a health hazard.

For this you should request the “

Decontamination Clearance Cer-

tificate

” from 

IKA

®

, or use the download printout of it from the 

IKA

®

 website 

www.ika.com

.

Return the instrument in its original packaging. Storage packaging is 
not sufficient. Also, please use suitable shipping package materials.

• Stirring bars: ø 6 mm, length up to 15 mm
   

 

ø 7 mm, length up to 60 mm

   

 

ø 10 mm, length up to 80 mm

• RS 1 

Set of magnetic stirring bars

• RSE 

Stirring bar remover

• H 15 

Bath attachment 

• H 28 

Bath attachment

• H 29 

Oil bath attachment 

• H 30 

Oil bath attachment

• H 16 V  Support rod 

• H 16.3  Extension

• H 38 

Holding rod 

• H 44 

Cross sleeve

• ETS-D5  Contact thermometer
• ETS-D6  Contact thermometer

 Accessories

Summary of Contents for RCT basic

Page 1: ...20000005393 RCT basic_072017 RCT basic safety control...

Page 2: ...cemakers Risk of burns Exercise caution when touching the housing parts and the heating plate The heating plate can reach temperatu res in excess of 310 C Pay attention to the residual heat after swit...

Page 3: ...of the assembly must lie within the surface on which it is set up Always disconnect the plug before tting accessories The appliance can only be disconnected from the mains supply by pulling out the m...

Page 4: ...he manufacturer or if the appliance is operated improperly contrary to the manufacturer s speci cations or if the appliance or the printed circuit board are modi ed by third parties Correct use Assemb...

Page 5: ...heating function by pressing the rotary knob C also on use of a contact thermometer The set point and actual temperatures will be shown alternately on the display E L D When the heating is switched o...

Page 6: ...when the device is started up Change the mode Move device switch A to the OFF position Press and hold rotary knobs C and D Move device switch A to the ON position Release rotating knobs C and D D The...

Page 7: ...or anticlo ckwise stop This will cause irreparable damage to the potentio meter Set the main switch to the ON position Using the screwdriver supplied turn the Safe Temp setting screw to the clockwise...

Page 8: ...Contact thermometer without safety circuit acc to DIN 12 878 class 0 The instrument only heats if the test current circuit is closed by an electrical connection of the plug pins 3 and 5 2 wire connec...

Page 9: ...g surfactant Cosmetics isopropyl alcohol water containing surfactant Foodstuffs water containing surfactant Fuels water containing surfactant Do not allow moisture to get into the appliance when clean...

Page 10: ...nsor of the base plate is higher than the set safe temperature limit Heating off Switch off device until the surface temperature of the base plate is lower than the selected safe temperature limit Set...

Page 11: ...missible relative humidity 80 Protection type acc DIN EN 60529 IP 42 Protection class I Overvoltage category II Contamination level 2 Fuse F1 F2 T6 3A 250V at nominal voltage 230 V T10A 250V at nomina...

Page 12: ...Base plate p y Heating y Adjustable safety circuit j y The control accuracy values given were determined using the following set up 500 ml water in 600 ml glass beaker magnetic stirring bar 40 600 rp...

Reviews: