32
› The device is maintenance-free� It is only subject to the natural wear and tear of components
and their statistical failure rate�
/// Cleaning:
› For cleaning disconnect the mains plug!
› Use only cleaning agents which have been approved by IKA to clean IKA devices�
These are water (with tenside) and isopropanol�
› Wear protective gloves during cleaning the devices�
› Electrical devices may not be placed in the cleansing agent for the purpose of cleaning�
› Do not allow moisture to get into the device when cleaning�
› Before using another than the recommended method for cleaning or decontamination, the
user must ascertain with IKA that this method does not destroy the device�
/// Spare parts order:
› When ordering spare parts, please give:
- device type�
- serial number, see type plate�
- software version (upon device starting up)�
/// Repair:
Please only send devices in for repair that have been cleaned and are free of materials which
might present health hazards�
If your appliance requires repair, return it in its original packaging� Storage packaging is not
sufficient when sending the device - also use appropriate transport packaging�
Maintenance and cleaning