IKA NANOSTAR 7.5 digital Manual Download Page 5

8

9

/// Working with the device

 

Danger!  

›  Do not use the device in explosive atmospheres, it is not EX-protected� 
›  With substances capable of forming an explosive mixture, appropriate safety measures must 

be applied, e�g� working under a fume hood�

›  To avoid body injury and property damage, observe the relevant safety and accident preven-

tion measures when processing hazardous materials�

›  The device is not suitable for hand-held operation�
›  Do not touch rotating parts during operation!

 

Warning!    

›  Pathogenic materials must only be processed in closed vessels under a suitable  fume hood� 

If you have any questions, please contact IKA application support�

›  The operation of a free rotating shaft end is prohibited� Therefore, for safety reason, only 

insert through the stirring tool over the upper edge of housing at standstill�

 Caution!

›  Make certain that the unit is set at the lowest speed before commissioning; otherwise, the 

unit will begin running at the speed set in last operation� Gradually increase the speed� 
Please observe the max� permitted speed for the stirring element� Never set higher speed�

!

 Notice!

›  Covers or parts that can be removed from the device without tools must later be refitted to 

ensure safe operation� This will prevent the infiltration of foreign objects, liquids and other 
contaminants�

›  Imbalance of the output shaft, the chuck and in particular the stirring tools can lead to 

uncontrolled resonant vibrational behavior of the device and the whole assembly� Glass ap-
paratus and stirrer containers can be damaged or shattered by this� It can cause injury to the 
operator, also can damage the rotating stirring tool� In this case exchange the stirring tool 
for one without imbalance or remedy the cause of the imbalance� If there is still imbalance, 
return it to the dealer or the manufacturer along with a description of the fault�

›  There may be electrostatic activity between the medium and the output shaft which could 

cause a direct danger�

›  Never operate the device with the stirrer tools rotating freely� Ensure that parts of the body, 

hair, jewelry or items of clothing cannot be trapped by the rotating parts�

›  Pay attention when setting the speed to any imbalance of the stirrer tools and possible 

spraying of the medium to be stirred�

›  Ensure that the stand does not start to move�
›  Only process media that will not react dangerously to the extra energy produced through 

processing� This also applies to any extra energy produced in other ways, e�g� through light 
irradiation�

›  Reduce the speed if:

- the medium splashes out of the vessel because the speed is too high,
- the device is not running smoothly,
- the device begins to move around because of dynamic forces,
- an error occurs�

/// Accessories

›  Protect the device and accessories from bumping and impacting�
›  Check the device and accessories beforehand for damage each time when you use them� Do 

not use damaged components�

›  Safe operation is only guaranteed with the accessories described in the ”Accessories” section�
›  Must disconnect the power before changing stirring element and fitting allowed accessories�

/// Power supply / switching off the device

!

 Notice!

›  If the device is operated too long in overload or if the ambient temperature is too high, the 

device switches off permanently�

›  Disconnecting the device from the power outlet only works by removing the main cord from 

the socket� 

›  The device must only be operated with the original power supply unit�
›  The power supply comes with interchangeable plug adapters for different countries so make 

sure the appropriate adapter is used�

›  The outlet for the power supply must be easily accessible�
›  After an interruption in the power supply during a stirring process, the device does not 

restart automatically� 

›  The device must only be opened by trained technicians, even during repair� The device must 

be unplugged from the power supply before opening� Live parts inside the device may still 
be live for some time after unplugging from the power supply�

/// Disposal instructions

›  The device, accessories and packaging must be disposed of in accordance with local and 

national regulations�

Intended use

///

 

Use

›  For mixing / stirring liquids from low to medium viscosities with various stirring tools�

› 

Stand device

 (chuck points down)

/// Range of use 

Indoor environments similar to that a laboratory of research, teaching, trade or industry�

The safety of the user cannot be guaranteed:
›  if the device is operated with accessories that are not supplied or recommended by the 

manufacturer�

›  if the device is operated improperly or contrary to the manufacture’s specifications�
›  if the device or the printed circuit board are modified by third parties�

Summary of Contents for NANOSTAR 7.5 digital

Page 1: ...IKA NANOSTAR 7 5 digital...

Page 2: ...3 ENGLISH...

Page 3: ...of warning symbols 6 Safety instructions 7 Intended use 9 Unpacking 10 Technical data 23 Operation 18 Speed 11 Output shaft 11 Operator panel and display 12 USB interface 19 Maintenance and cleaning 2...

Page 4: ...e Place the power supply outside the working area of the device Avoid knocking and impacting on the lower end of the shaft and the chuck gear teeth Even minor invisible damage can lead to imbalance an...

Page 5: ...spraying of the medium to be stirred Ensure that the stand does not start to move Only process media that will not react dangerously to the extra energy produced through processing This also applies...

Page 6: ...rmal operations the speed value on the display corresponds to the speed of the output shaft in revolutions per minute rpm Output shaft The clamping chuck and output shaft permit clamping IKA recommend...

Page 7: ...sconnection from the power supply network B Display Display the speed value C Control knob Rotating Pressing Set the speed by rotating Start stop the stirrer function by pressing Installation Assembli...

Page 8: ...14 15 2 x 1 2 rpm Assembling the stirring element using the chuck Assembling the stirring shaft protector rpm 2 3 1 Securing the mixing vessel to the stand 2 1 2 2 1 1 2 1 1 2 3...

Page 9: ...eable adapter Notice The AC adapter is suitable for 100 VAC to 240 VAC The exchangeable adapter for the relevant country must be attached before the AC adapter is plugged into the mains socket Click C...

Page 10: ...ne is ready for operation when the mains plug is plugged in If these conditions are not met safe operation is not guaranteed and the machine could be damaged Observe the ambient conditions temperature...

Page 11: ...nents and their statistical failure rate Cleaning For cleaning disconnect the mains plug Use only cleaning agents which have been approved by IKA to clean the device water containing surfactant isopro...

Page 12: ...oper use insufficient care or maintenance not carried out in accordance with the instructions in this operating manual Technical data Drive unit Speed range under nominal load rpm 0 50 2000 Speed adju...

Page 13: ...England LTD Phone 44 1865 986 162 eMail sales england ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ik...