IKA I-MAG 300 Manual Download Page 6

10

11

Installation

Useful information

!

 Notice!

›  The socket used must be earthed (fitted with earth contact)�
›  Observe the ambient conditions (temperature, humidity, etc�) listed under “Technical data”�

/// Connect the I-MAG 300 to the I-MAG CP

/// Place the vessel with a magnetic rod

/// Connect the I-MAG CP to the power supply

The magnetic stirrer is suitable for continuous operation� The motor current is limited electron-
ically� The device has an anti-stall and anti-overload system� If a fault occurs, a safety circuit 
immediately switches off the motor permanently�

/// Speed – normal operation

Speed - regulated (no variation in speed):

The speed is monitored and regulated by processor controlled� The target speed is constantly 
compared with the actual speed value� This guarantees a constant speed even if the viscosity of 
the substance being stirred changes�
Fluctuations in mains voltage within the permitted tolerance range have no effect on the quality 
of regulation and constancy of speed� The speed is set using the rotating / pressing knob� Dur-
ing normal operations the speed value on the display corresponds to the speed of the stirring 
bar in revolutions per minute (rpm)�

/// Speed – overload operation

When adding solid or semi-viscous substances, the magnetic stirrer can be operated at 130 % 
of its rated power for a short time�
In overload operation, e�g� due to the increasing viscosity of the medium to be stirred, the speed 
is reduced to such an extent that the rated current of the unit is not exceeded�

Overload status:

The device is already running in overload mode if the actual speed does not match the setpoint 
speed� 
If the motor current or the device's internal temperature exceeds the permissible limit, you will 
be informed by a warning message in the display�

!

 Notice!

›  Small vessels, especially those made of metal, must be secured against twisting!

1.

2.

2.1

I-MAG CP

I-MAG 300

2.2

Click

1.

OFF

3.

100 - 240 V ~
50/60 Hz
140 W
 IP 22

2.

Menu

Back

1.

2.

Summary of Contents for I-MAG 300

Page 1: ...I MAG...

Page 2: ...ENGLISH...

Page 3: ...ation of conformity 6 Explication of warning symbols 6 Interfaces and outputs 22 Operator panel and display 12 Warranty 27 Useful information 10 Maintenance and cleaning 24 Accessories 24 Intended use...

Page 4: ...hazardous situation which if not avoided can result in death serious injury Warning Indicates a potentially hazardous situation which if not avoided can result in injury Caution Indicates practices wh...

Page 5: ...l IKA accessories Always disconnect the plug before attaching accessories Observe the operating instructions of the accessories Power supply Switching off the device Warning The device will automatica...

Page 6: ...ge within the permitted tolerance range have no effect on the quality of regulation and constancy of speed The speed is set using the rotating pressing knob Dur ing normal operations the speed value o...

Page 7: ...terface analog signal output O Setup plate P Connection cable I MAG CP G A H I J K L M N Menu O P C B D E F Back Display 1 Set speed value 2 Actual speed value 3 Counter Timer 4 Button lock 5 Pedal sw...

Page 8: ...00 500 247 actual rpm set rpm Counter A dd hh mm 18 ss 00 00 00 500 500 actual rpm set rpm 1 3 5 2 4 Back Back Counter A dd hh mm 28 ss 00 00 00 500 0 actual rpm set rpm Counter A dd hh mm 00 ss 00 00...

Page 9: ...m etc Stirring bar decoupling monitoring does not work if the stirring bar already breaks away before the system has reached the set target speed Intermittent mode The Intermittent mode is a cyclic in...

Page 10: ...nob Using the Edit Insert or Delete buttons you can make the corresponding changes In a program user can define up to 10 program segments The selected program segment is highlighted in blue When editi...

Page 11: ...onding parameter you want to reset with the rotating pressing knob The system will prompt you to confirm the restoration of the selected factory setting Press OK button to reset all the system setting...

Page 12: ...and their properties are detected automatically The USB interface is used in conjunction with software control for remote operation and can also be used for software update of the device USB device dr...

Page 13: ...tear of components and their statistical failure rate Cleaning For cleaning disconnect the mains plug Use only cleaning agents which have been approved by IKA to clean IKA devices These are water cont...

Page 14: ...from improper use insufficient care or maintenance not carried out in accordance with the instructions in this operating manual Technical data General data I MAG 300 I MAG CP Voltage 100 240 V Frequen...

Page 15: ...2 eMail sales england ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ika cn POLAND IKA Poland Sp z o o P...