IKA I-MAG 300 Manual Download Page 14

26

27

Speed display set-value

-

TFT

Speed display actual-value

-

TFT

Speed control

-

rotating / pressing knob

Speed range

100 ��� 1500 rpm

-

Setting accuracy speed 

10 rpm

-

Speed deviation (no overload)

± 10 rpm

-

Stirring bar length

30 ��� 140 mm

-

Self-heating of the setup plate by max� 
stirring (RT: 22 °C / duration: 1 h)

+ 26 K

-

Timer function

Timer

-

yes

Timer display

-

TFT

Timer control

-

b rotating / pressing 

knob

Time setting range

-

0 / 1 minutes ��� 99 days + 23 

hours + 59 minutes

Analog signal

Analog input:

Measuring voltage range
Input resistance 

0 ��� +5 V
> 100 kΩ

Analog output:

Operating voltage range 
Output loadability

0 ��� +5 V 

> 2 kΩ

 Subject to technical changes!

Warranty

›  In accordance with IKA warranty conditions, the warranty period is 24 months� For claims 

under the warranty please contact your local dealer� You may also send the device direct to 
our factory, enclosing the delivery invoice and giving reasons for the claim� You will be liable 
for freight costs�

›  The warranty does not cover worn out parts, nor does it apply to faults resulting from 

improper use, insufficient care or maintenance not carried out in accordance with the 
instructions in this operating manual�

Technical data

General data

I-MAG 300

I-MAG CP

Voltage

-

100 ��� 240 V

Frequency

-

50 / 60 Hz

Power input

140 W

Power input standby 

-

5 W

DC voltage

48 V

-

Motor rating output 

90 W

-

USB interface 

-

yes

RS 232 interface 

-

yes

Bluetooth interface 

-

no

Interface analog signal

-

yes

Interface (Pedal switch)

-

yes

Ethernet interface 

-

yes

Programs

-

yes

Firmware update

-

yes

Wireless control

-

no

Automatic reverse rotation 

yes

-

Intermittent mode

yes

-

Break detection stirring bar

yes

Permissible ambient temperature

+  5 ��� +  40 °C

Permissible relative humidity

80 %

Protection class according to 
DIN EN 60529

IP 64

IP 22

Material of installation surface

stainless steel 1�4301

-

Dimensions of installation area

260 x 260 mm

-

Dimensions (W x D x H)

260 x 300 x 70 mm

335 x 165 x 65 mm

Weight

6 kg

2 kg

Protection class

I

Contamination level

2

Overvoltage category

II

Operation at a terrestrial altitude

max� 2000 m

Stirring function

Number of stirring positions 

1

-

Stirring quantity max� per stirring position (H

2

O)          

with I-MAG SP 150
with I-MAG SP 300

150 l 
300 l

-

Max� bearing load, I-MAG 300

75 kg

-

Direction of rotation

right / left

-

Summary of Contents for I-MAG 300

Page 1: ...I MAG...

Page 2: ...ENGLISH...

Page 3: ...ation of conformity 6 Explication of warning symbols 6 Interfaces and outputs 22 Operator panel and display 12 Warranty 27 Useful information 10 Maintenance and cleaning 24 Accessories 24 Intended use...

Page 4: ...hazardous situation which if not avoided can result in death serious injury Warning Indicates a potentially hazardous situation which if not avoided can result in injury Caution Indicates practices wh...

Page 5: ...l IKA accessories Always disconnect the plug before attaching accessories Observe the operating instructions of the accessories Power supply Switching off the device Warning The device will automatica...

Page 6: ...ge within the permitted tolerance range have no effect on the quality of regulation and constancy of speed The speed is set using the rotating pressing knob Dur ing normal operations the speed value o...

Page 7: ...terface analog signal output O Setup plate P Connection cable I MAG CP G A H I J K L M N Menu O P C B D E F Back Display 1 Set speed value 2 Actual speed value 3 Counter Timer 4 Button lock 5 Pedal sw...

Page 8: ...00 500 247 actual rpm set rpm Counter A dd hh mm 18 ss 00 00 00 500 500 actual rpm set rpm 1 3 5 2 4 Back Back Counter A dd hh mm 28 ss 00 00 00 500 0 actual rpm set rpm Counter A dd hh mm 00 ss 00 00...

Page 9: ...m etc Stirring bar decoupling monitoring does not work if the stirring bar already breaks away before the system has reached the set target speed Intermittent mode The Intermittent mode is a cyclic in...

Page 10: ...nob Using the Edit Insert or Delete buttons you can make the corresponding changes In a program user can define up to 10 program segments The selected program segment is highlighted in blue When editi...

Page 11: ...onding parameter you want to reset with the rotating pressing knob The system will prompt you to confirm the restoration of the selected factory setting Press OK button to reset all the system setting...

Page 12: ...and their properties are detected automatically The USB interface is used in conjunction with software control for remote operation and can also be used for software update of the device USB device dr...

Page 13: ...tear of components and their statistical failure rate Cleaning For cleaning disconnect the mains plug Use only cleaning agents which have been approved by IKA to clean IKA devices These are water cont...

Page 14: ...from improper use insufficient care or maintenance not carried out in accordance with the instructions in this operating manual Technical data General data I MAG 300 I MAG CP Voltage 100 240 V Frequen...

Page 15: ...2 eMail sales england ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ika cn POLAND IKA Poland Sp z o o P...