![IKA G-L Manual Download Page 8](http://html1.mh-extra.com/html/ika/g-l/g-l_manual_624627008.webp)
14
15
Operation
RCF (g-force) calculation in accordance with DIN 58970:
Rcf = 1.118 * 10¯⁵ * n² * r
max
n:
rotational speed in rpm
r
max
:
maximum centrifugation radius in cm
Example, The maximum centrifugation radius of the rotor IKA G-L CR 12x2ml is 6 cm� At a
speed of 15700 rpm, a maximum g-force of 16500 x g can be achieved�
Switching between rcf and rpm / Setting rcf or rpm
Switching between counter and timer / Setting timer
1
rcf rpm
...
2
1
counter timer
...
2
Summary of Contents for G-L
Page 1: ...IKA G L...
Page 2: ...2 3 ENGLISH...
Page 10: ...18 19 Editing the program 1 2 3 1 4 2 3 Activating the program 1 2 3...