Safety
instructions
11
Reduce the stirring speed, if
› the medium splashes out of the vessel because the speed is too high.
› the device is not running smoothly.
› the container moves on the stirring plate.
› an error occurs.
/// Accessories
› Safe operation is only guaranteed with the accessories described in the
"Accessories" chapter.
› Always disconnect the plug before attaching or removing accessories.
› Observe the operating instructions of the accessories.
› Ensure that the external temperature sensor is inserted into the medium to
a depth of at least 20 mm when connected.
› Accessories must be securely attached to the device and must not come
off by themselves. The center of gravity of the assembly must lie within the
surface on which it is set up.
› External temperature probes can react with the medium and cause a
short-circuit.
1. It is important to use a sensor that is isolated against electricity. Our
recommendation is to use temperature probe PT 1000.90 (xxxx). Please
ensure that the coating is resistant to the solvent being used and in an
undamaged condition.
2. A non-insulated or not fully coated temperature sensor can seriously
damage or destroy ElectraSyn 2.0. Measuring outside of the vial, e.g. inside
a dry block, a metal probe can be used.