C7010/12 0508
24
C14
C5010.4
Replacing the crucible holder
For the use of the oneway crucible C14
the decomposition vessels C 7010
and C 7012 must be put in the
in the foundation C5010.4.
These foundation C 5010.4
replaces the normal crucible holding.
Take the cover part and until the ignition
wire on the crucible holder like
discribed before.
Screw down the cru-
cible holder from the
shaft and replace it
with the foundation
C 5010.4.
Installation of the oneway crucible C14
The oneway crucible C14 must
be put into the crucible holding
of the foundation C 5010.4 so
that the ignition wire touches
the oneway crucible C14.
It would the best to stick it
under the ignition wire.
Operating pressure PS
max. 230 bar (30MPa)
Test pressure PT
330 bar (33 MPa)
Volume
210 ml
Nuclear temperatures
≥ 1300 °C
Operating pressures
max. 50°C
Sample decomposition
isochor
Oxygen valve
non-returnable valve
(Vacuum operation not possible)
Crucible holder
as in DIN 51 900
Detection temperature
Pt 100 integratet
Material
Lower part
3.4365
Threadet joint
3.4365
C7010
C7012
Cover
1.4439
2.4610
Insert
1.4439
2.4610
Tec hnic al dat a
3017500b_C7010_7012_052008:C5010_12 09.06.2008 9:59 Uhr Seite 24