The maintenance menu offers you the option of performing a series of system func-
tions manually in the event of a malfunction. The commands of the Maintenance
menu can only be performed if the measuring cell is in the Waiting mode.
The functions
2SHQ0&
,
&ORVH 0&
, and
2 ILOO
are activated by the corresponding
menu command and are automatically terminated. The menu item in question is
locked while the function is being performed.
2SHQ0&
Opens the measuring cell cover
&ORVH0&
Closes the measuring cell cover
&RROLQJZDWHU
Water flows through the outer vessel and the cooling water
temperature is displayed in a window. The check can be ter-
minated manually with
2.
or will be terminated automatically
after 5 minutes.
7HPSHUDWXUH
Water flows through the outer vessel and regulation is
(TXDOL]DWLRQ
activated. The current target value (25 or 30 °C) is used for
this process. The actual value of the temperature is displayed
in a window. To end the check, press
2.
.
6WLUUHURQRII
The stirrer is turned on and off.
(PSW\,9
Water is pumped out of the inner vessel.
Emptying is ended by time control (default: 90 sec).
7HPS,QLW
The temperature measurement is reinitialized.