Close the decomposition vessel (C 5010, C 5012 or C 62).
Place the cover onto the lower section and push down until it presses against the
stop piece in the lower section.
C 5010
C 5012
C 62
Place the union nut onto the lower section and tighten by hand.
The decomposition vessel can now be filled.
3HUIRUPLQJWKHPHDVXUHPHQW
c
Guide the decomposition vessel into the filler head of the open measuring cell cover
until it catches in place.
$OZD\VKROGWKHGHFRPSRVLWLRQYHVVHOVHFXUHO\RQWKHWRSE\WKHFDSVFUHZ
The decomposition vessel maintains a defined position in the holder on the fill head
by being lowered at the centre of the fill head by 0.8 mm. A spring element then
makes contact with the electrical ignition contact on the decomposition vessel.
7KHGHFRPSRVLWLRQYHVVHOLVQRZVXVSHQGHGYHUWLFDOO\LQWKHUHFHLYLQJDUHD
FKHFNHYHU\WKLQJRQHODVWWLPH