23
Error codes
Accessories
A10.1
Standard beater
A10.3
Hard metal cutter
Warranty
In accordance with
IKA
®
warranty conditions, the war-
ranty period is 24 months. For claims under the warranty
please contact your local dealer. You may also send the
machine direct to our factory, enclosing the delivery in-
voice and giving reasons for the claim. You will be liable
for freight costs.
The fault is shown by an error code in the LED display (
C
) as following if the error occurs.
Proceed as follows in such cases:
F
Switch off the device using the main switch
F
Carry out corrective measures
F
Restart the instrument
Error code Cause
Effect
Solution
– – –
The cap is not closed properly
instrument
does not start
- Close the cover properly
E3
Temperature inside instrument is too high motor off
- Switch off the instrument and allow it to
cool down
E4
Motor blockage or overload
motor off
- Switch off the instrument
- Decrease the motor’s load and reduce the
material to be ground
E25
Safety relay fault
motor off
- Send the instrument for repair
E86
Voltage is too low
motor off
- Switch off the instrument
- Check the Mains voltage
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.
A10.2
Star-shaped knife
A10.4
Grinding chamber reduction
The warranty does not cover worn out parts, nor does
it apply to faults resulting from improper use, insufficient
care or maintenance not carried out in accordance with
the instructions in this operating manual.
Summary of Contents for A10 basic