ISO81060-2

2013,(Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of

Automated Measurement Type).

ELECTROMAGNETIC COMPATIBILITY INFORMATION

This product is applicable to the equipment and system requirements for the purpose of
receiving radio frequency energy for the purpose of the work, Bluetooth receive bandwidth
2M. This product can also be used to include RF transmitter equipment and system
requirements and emission frequency of 2.4GHz ISM band, Bluetooth modulation
types:GFSK, effective

max 

radiated power: <

4

dBm

Table 1 - Emission

Phenomenon

Compliance

Electromagnetic environment

RF emissions

CISPR 11

Group 1, Class B

Home healthcare environment

Harmonic distortion

IEC 61000-3-2

Class A

Home healthcare environment

Voltage

fluctuations

and flicker

IEC 61000-3-3

Compliance

Home healthcare environment

Table 2 - Enclosure Port

Phenomenon

Basic EMC standard

Immunity test levels

Home Healthcare Environment

Electrostatic
Discharge

IEC 61000-4-2

±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air

Radiated RF EM field

IEC 61000-4-3

10V/m
80MHz-2.7GHz
80% AM at 1kHz

Proximity fields from
RF

wireless

communications
equipment

IEC 61000-4-3

Refer to table 3

Rated

power

frequency

magnetic

fields

IEC 61000-4-8

30A/m
50Hz or 60Hz

Table 3 – Proximity fields from RF wireless communications

equipment

Test frequency
(MHz)

Band
(MHz)

Immunity test levels

Professional healthcare facility environment

385

380-390

Pulse modulation 18Hz, 27V/m

450

430-470

FM,

±

5kHz deviation, 1kHz sine, 28V/m

710

704-787

Pulse modulation 217Hz, 9V/m

745
780

Summary of Contents for Neo BP5S

Page 1: ...DICATORS 2 SET UP REQUIREMENTS 3 SET UP PROCEDURES 3 MEASUREMENT PROCEDURES 3 TAKING YOUR BLOOD PRESSURE READING 5 SPECIFICATIONS 6 GENERAL SAFETY AND PRECAUTIONS 6 BATTERY HANDLING AND USAGE 8 TROUBL...

Page 2: ...sure measurement system It is designed to measure the systolic and diastolic blood pressures and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around th...

Page 3: ...e iHealth Cloud Account Your iHealth account also gives you access to the free and secure iHealth cloud service Go to www ihealthlabs com and sign in with the same account Power on the Monitor Press S...

Page 4: ...s measured position the monitor in the middle of your arm so that it is aligned with your middle finger When the right arm is measured position the monitor in the middle of your arm so that it is alig...

Page 5: ...Note Please consult a health care professional for interpretation of pressure measurements BLUETOOTH FUNCTION Connect to iOS Device via Bluetooth a Enable Bluetooth on your iOS device b Launch the iH...

Page 6: ...r storage and transport 90 RH 17 Environmental pressure 80kPa 105kPa 18 Battery life more than 130 measurements on a full charge 19 The blood pressure measurement system includes accessories pump valv...

Page 7: ...that complies with IEC60601 1 UL 60601 1 and IEC 60601 1 2 EN 60601 1 2 such as UE05WCP 050100SPC input 100 240V 50 60Hz 0 18A output DC 5V 1 0A Please note that the monitor jack size is USB microAB...

Page 8: ...tery low charge symbol flashing for a few seconds Do not change the battery If the battery can no longer be charged please contact Customer Service Overcharging the battery may reduce its lifetime Lit...

Page 9: ...Bluetooth communicate error Reset monitor by pressing the START STOP button and holding for about 10 seconds then connect the mobile device correctly and try again If the monitor is still abnormal ple...

Page 10: ...ed with Ethyl alcohol 75 90 Then air dry the cuff 8 It is recommended that product performance be checked every 2 years or after each repair Please contact the customer service Do not drop this monito...

Page 11: ...e equipment Note This product has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable prot...

Page 12: ...tre antenne ou metteur CE Statement The CE Mark applies to products regulated by certain European health safety and environmental protection legislation The CE Mark is obligatory for products it appli...

Page 13: ...ent Harmonic distortion IEC 61000 3 2 Class A Home healthcare environment Voltage fluctuations and flicker IEC 61000 3 3 Compliance Home healthcare environment Table 2 Enclosure Port Phenomenon Basic...

Page 14: ...nts burst IEC 61000 4 4 2 kV 100kHz repetition frequency Surges Line to line IEC 61000 4 5 0 5 kV 1 kV Surges Line to ground IEC 61000 4 5 0 5 kV 1 kV 2 kV Conducted disturbances induced by RF fields...

Reviews: