iHealth CardioMed User Manual Download Page 15

 

15

 

/

 

16

 

 

ISO81060-2 

2013,(Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement 

Type). 

ELECTROMAGNETIC COMPATIBILITY INFORMATION 

This product is applicable to the equipment and system requirements for the purpose of receiving radio frequency 

energy for the purpose of the work, Bluetooth receive bandwidth 2M. This product can also be used to include RF 

transmitter equipment and system requirements and emission frequency of 2.4GHz ISM band, Bluetooth modulation 

types:GFSK, effective radiated power: <4dBm.

 

Table 1 - Emission 

Phenomenon Compliance 

Electromagnetic environment 

RF emissions 
 

CISPR 11 

Group 1, Class B 

Home healthcare environment 

Harmonic distortion 

IEC 61000-3-2 

Class A 

Home healthcare environment 

Voltage fluctuations 
and flicker 

IEC 61000-3-3 

Compliance 

Home healthcare environment 

 

 

 

Table 2 - Enclosure Port

 

Phenomenon Basic EMC 

standard 

Immunity test levels 

Home Healthcare Environment 

Electrostatic 
Discharge 

IEC 61000-4-2 
 

±8 kV contact 
±2kV, ±4kV, ±8kV, ±15kV air 

Radiated RF EM 
field 
 

IEC 61000-4-3 
 

10V/m 
80MHz-2.7GHz 
80% AM at 1kHz 

Proximity fields from 
RF wireless 
communications 
equipment 

IEC 61000-4-3 
 

Refer to table 3 

Rated power 
frequency magnetic 
fields  

IEC 61000-4-8 

30A/m 
50Hz or 60Hz 

 

Table 3 –    Proximity fields from RF wireless communications 

equipment 

 

Test frequency   
(MHz) 

Band 
(MHz) 

Immunity test levels 

Professional healthcare facility environment 

385 

380-390 

Pulse modulation 18Hz, 27V/m 

450 

430-470 

FM, ±5kHz deviation, 1kHz sine, 28V/m 

710 

704-787 

Pulse modulation 217Hz, 9V/m 

745 
780 
810 

800-960 

Pulse modulation 18Hz, 28V/m 

Summary of Contents for CardioMed

Page 1: ...NTS 6 BATTERY HANDLING AND USAGE 6 SET UP PROCEDURES 7 ABPM SYNC AND SETUP 7 MEASUREMENT PROCEDURES 8 TAKING YOUR BLOOD PRESSURE READING 9 TAKING YOUR BLOOD PRESSURE READING MANUALLY 9 TAKING YOUR BLO...

Page 2: ...hypertension z Assessment of treatment Assessing 24 h BP control Identifying true resistant hypertension z Assessing morning hypertension and morning BP surge z Screening and follow up of obstructive...

Page 3: ...ce of adjacent pulse period is 0 14s and more than 53 percent of the total number of pulses readings falls within this definition z This product should not be used as a USB device z If the blood press...

Page 4: ...IBILITY INFORMATION It is suggested that the blood pressure monitor be kept 10 meters away from other wireless devices such as WLAN unit microwave oven etc z The results of blood pressure measurements...

Page 5: ...l of the cuff Please consult a physician for specific situation z It is recommended that the patient record important events such as take medicine taking time drug name drug dose etc awake and asleep...

Page 6: ...asurements on a full charge 19 The blood pressure measurement system includes accessories pump valve cuff LED screen and sensor Note These specifications are subject to change without notice SET UP RE...

Page 7: ...Health service provider Battery life and charge cycles vary by use and settings SET UP PROCEDURES Download the Free iHealth App Prior to first use scan QR code and download the iHealth App Follow the...

Page 8: ...surement Lying Down During Measurement a Lie on your back b Place your arm straight along your side with your hand palm side up c The cuff should be placed at the same level as your heart Note Blood p...

Page 9: ...tep to achieving a successful test TAKING YOUR BLOOD PRESSURE READING MANUALLY a Long press the button the monitor will activate and all display characters are shown for self test You can check the LE...

Page 10: ...d When Auto Mode on the monitor will continue take measurements in the scheduled time and after the monitor get enough readings the monitor automatic stop taking readings Note Please consult a health...

Page 11: ...g the button and holding for about 10 seconds then connect the mobile device correctly and try again If the monitor is still abnormal please contact the local distributor or the factory LED display re...

Page 12: ...2 years or after each repair Please contact the customer service z Do not use the device if you think it is damaged or if anything appears unusual Please refer your device to qualified service personn...

Page 13: ...Paris France support ihealthlabs eu www ihealthlabs eu ANDON HEALTH CO LTD No 3 Jinping Street YaAn Road Nankai District Tianjin 300190 China Tel 86 22 60526161 IMPORTANT INFORMATION REQUIRED BY THE...

Page 14: ...pe and its gain should be so chosen that the equivalent isotropically radiated power e i r p is not more than that necessary for successful communication Conform ment la r glementation d Industrie Can...

Page 15: ...n IEC 61000 3 2 Class A Home healthcare environment Voltage fluctuations and flicker IEC 61000 3 3 Compliance Home healthcare environment Table 2 Enclosure Port Phenomenon Basic EMC standard Immunity...

Page 16: ...4 2 kV 100kHz repetition frequency Surges Line to line IEC 61000 4 5 0 5 kV 1 kV Surges Line to ground IEC 61000 4 5 0 5 kV 1 kV 2 kV Conducted disturbances induced by RF fields IEC 61000 4 6 3V 0 15...

Reviews: