It can help to prevent infection, the potential spread of infection, and cross-
contamination. The cleaning and disinfection is absolutely necessary for the test 
procedure, because cleaning can ensure the meter works well (for example, the 
display will be clear to see after cleaning). 
Disinfection can avoid the infection to you or to the other people, and the cross-
infection. Cleaning removes dirt and debris but does not eliminate bacteria or virus, 
while disinfection is to eliminate bacteria or virus on the meter. 
Clean and disinfect the meter and lancing device each time before the 
measurement. 
Cleaning is to wipe the whole surface of the meter for 10 seconds, until there is no 
soil on the surface. 
Disinfection is to wipe the whole surface of the meter with a gentle disinfecting 
solution for 2 min, and keep the surface wet during the 2min. 

The iHealth Wireless Smart Glucose Meter and the iHealth lancing device should 
be cleaned and disinfected at a minimum of once per week.  

The iHealth Wireless Smart Glucose Meter and the iHealth lancing device are 
validated to support 10,000 individual tests—and consequently 10,000 cleanings 
over their 5 years life spans. The meter and the lancing device are for single-patient 
use. If the user tests 6 times every day, the meter and lancing device should be 
cleaned and disinfected 6 times per day which equals 10950 cycles over the 5 year 
life of the devices. The meter and lancing device were validated for 11,000 cycles 
which could support up to 6 cleaning and disinfection cycles per day. 
If the meter is being operated by a second person who is providing testing 
assistance to the user, the meter and lancing device should be disinfected prior to 
use by the second person. 

Find below, how to clean the meter and lancing device. 
1. After a test, wash your hands thoroughly with soap and water after handling the
meter, lancing device, or test strips. 
2. Use the clean wipe to carefully clean the meter, front and back.
3. Then, disinfect the meter with another wipe and allow the surface to dry naturally:
the meter should remain wet for 2 minutes.  
4. Use the same method with the clean wipes to clean and disinfect the lancing
device. 

Note: 

1

 

 

You should always perform  a cle aning step prior the a disinfectio n 

step. Wash hands thoroughly with soap and water after handling the meter, lancing 

device, or test strips. 

2

Only the surface of the mete r can be cleaned and disinfected with 

the disinfecting wipe. Do not insert the disinfecting wipe into the test strip port.  

3

Avoid getting moistur e inside the  meter through the strip port at all 

times.   

4

If you have any questions you can call the number: 1-855-816-7705 

Summary of Contents for BG1S

Page 1: ...iHealth Wireless Smart Gluco Monitoring System BG1S OWNER S MANUAL For in vitro diagnostic use only Read instructions before use for self testing...

Page 2: ...9 About the online test 9 CLEANING AND DISINFECTION 12 INFORMATION ABOUT ALTERNATE SITE TESTING AST 14 What is Alternate Site Testing 14 What is the advantage of Alternate Site Testing 14 When should...

Page 3: ...at home as an aid to monitor the effectiveness of diabetes control The iHealth Wireless Smart Gluco Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal u...

Page 4: ...her bloodborne pathogens For over the counter use Important safety instructions Please read the following information carefully before using the iHealth Wireless Smart Gluco Monitoring System BG1S Alw...

Page 5: ...Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne Pathogens 2010 http www cdc gov injectionsafety Fingerstick DevicesBGM html YOUR NEW WIRELESS SMART GLUCOSE METER Con...

Page 6: ...m direct sunlight Test strips must be stored in their original vial only Do not transfer them to a new vial or another container Do not touch the test strips when your hands are wet Use each strip pro...

Page 7: ...5 iHealth Lancing Device Use only with the iHealth lancet Use only iHealth Control Solution Keep the control solution at a temperature between 2 C and 30 C 36 F 86 F and at a humidity rate of 10 to 85...

Page 8: ...rt Gluco Monitoring System BG1S The iHealth Control Solution contains a specific quantity of glucose concentrate which reacts with the test strips it s used to verify the accuracy of blood glucose tes...

Page 9: ...contact with the control solution Once the test strip confirmation window is completely filled the meter will start counting down then you can remove the test strip from the control solution sample No...

Page 10: ...lycemic hyperosmolar state with or without ketosis Critically ill patients should not be tested with iHealth Wireless Smart Gluco Monitoring System BG1S If your blood glucose results are lower or high...

Page 11: ...to perform a blood glucose test When you perform a blood glucose test your results will be synchronized automatically on your iHealth profile WARNING When you get a new meter and you want to finish a...

Page 12: ...he battery according to following instructions before using it again 1 Turn the meter over with the iHealth logo facing down 2 Insert the narrow end of the pry plate into the machine groove close to t...

Page 13: ...and BG1S is nearby Step 4 d Follow app instructions to insert test strip Take a new test strip from the test strip vial hold the middle of the test strip with the arrow facing the meter and insert th...

Page 14: ...to compare your meter results with laboratory tests Refer to the instructions given by your physician or other qualified healthcare providers do not deviate from these instructions on the basis of the...

Page 15: ...se If the user tests 6 times every day the meter and lancing device should be cleaned and disinfected 6 times per day which equals 10950 cycles over the 5 year life of the devices The meter and lancin...

Page 16: ...Testing Pain is felt more readily on the fingertips because they are full of nerve endings receptors At other body sites where nerve endings are not so condensed pain is not felt as acutely When shou...

Page 17: ...below appear please contact iHealth Labs Customer Service Do not attempt to repair the meter by yourself and never try to disassemble the meter under any circumstances Troubleshooting Problems Possibl...

Page 18: ...Android or iOS mobile device can t begin testing when the meter is connected with the Android or iOS mobile device or when a test strip is inserted into the meter 2 Discoloration of the meter casing o...

Page 19: ...the limits for a C lass B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against har mful interference in a residential installation Th...

Page 20: ...ISPR 11 Group 1 Class B Home healthcare environment Harmonic distortion IEC 61000 3 2 Class A Home healthcare environment Voltage fluctuations and flicker IEC 61000 3 3 Compliance Home healthcare envi...

Page 21: ...riod of 3 years from the date of purchase by the original purchaser Warranty Period Under this Limited Warranty if a defect arises and a valid claim is received by iHealth within the Warranty Period r...

Page 22: ...Healthlabs eu Customer service https helpcenter ihealthlabs eu hc en gb requests new or menu Contact in App If you have questions or need assistance outside the operational hours and days please conta...

Page 23: ...y from direct sunlight or near heat sources Keep in a dry place Do not reuse Sterilized using irradiation Bluetooth sign Complies with the requirements of the European IVD Directive 98 79 EC This devi...

Page 24: ...FCC IC 20200819...

Reviews: