STERYLIS ULTRA Operating Manual
When the AUTO-TEST function is completed, the unit will go into standby mode.
Recommendations for maintenance
•
The device should be kept in a state of general cleanliness. Use only neutral detergents for cleaning.
•
Do not use for washing the water jet under pressure.
•
The air filters must be checked and replaced periodically as described in Chapter 7.2. The indication
of overuse / air filters need to be checked it is indicated on the user panel by displaying the
corresponding messages (see Chapter 9). The replacement of the filters must be carried out in
accordance with the instructions in Chapter 7.2.1 and 7.2.2. Do not lead to clogging of the filters or
for them being used excessively long.
•
After the service life of the UV lamps (9,000 operating hours) in the steriliser, they should be
replaced. This is related not only to the possible burning of the lamp but also to the decrease of its
maximum radiation dose over time. The UV lamps’ overrunning time is indicated on the user panel by
displaying an appropriate message (see Section 9). Only the qualified service technicians can replace
the lamp/s!
•
After the service life of the ozone generators (9,000 operating hours) in the steriliser, they should
be replaced. The ozone generators’ overrunning time is indicated on the user panel by displaying an
appropriate message (see Section 9). Only the qualified service technicians can replace the ozone
generators!
•
After the expiry date of the ozone sensor (12 months) or after 4,500 hours of operation, it must be
replaced by qualified service!
Maintenance
CAUTION!
CAUTION!
All maintenance work should only be carried out when the unit is disconnected
from the power supply (with the exception of alarm removal activities)!
All service activities (except filter replacement) may only be carried out by
authorized personnel and the manufacturer’s factory service. For details on
service, see the Warranty and Service section.