STERYLIS ULTRA Operating Manual
The safe operation of the product is also related to the principles listed below:
• The device may only be operated in sterilisation mode in a tightly closed room (no ozone may pene-
trate outside the sterilised room). Therefore, make sure that all windows, doors and any other escape
routes for ozone from the room are tightly closed. Gaps around and under doors and ventilation grilles
must be carefully sealed. The room must be protected from access by third parties.
• It is also forbidden to stay in rooms adjacent to the ozone room during the ozoning process. If the
ozoned room is not properly sealed, some ozone may penetrate into other rooms.
• To protect the room from access by third parties, it is essential to place warning signs “WARNING! NO
ENTRY! STERILISATION IN PROGRESS”. Place them in a clearly visible place. The signs are included in
the package.
• After starting the sterilisation mode, leave the room where the steriliser is operating as soon as
possible. The time to leave the room is 60 seconds from the start of the ozoning function, only after
that time the ozone generators start. It is forbidden to stay in the room during operation of the ozone
generator without a full respiratory and eye mask with a suitable absorber. It is recommended to use a
protective mask according to PN-EN 136 with an absorber according to PN-EN 14387, type NO P3. It is
also recommended to use personal protective equipment for the skin according to the table of personal
protective equipment.
• Smoking and the use of tools that cause a flame or spark is prohibited in an ozoned room.
• In order to deactivate the sterilisation mode before the end of the process, one should enter the room
wearing a full respiratory and eye mask with a suitable absorber (in accordance with EN 136 and EN
14387 standards) and in the skin protection equipment in accordance with the personal protective
equipment table.
• When the ozoning of the room is complete, it must be thoroughly ventilated for at least 30 minutes.
• Ozone is a heavier gas than air, after switching off the circulating fan of the steriliser, STERYLIS tends
to settle in the recesses of the ground and migrate by the floor.
• It is not recommended to sterilise rooms below ground level without mechanical ventilation.
• It is forbidden to carry out the process of ozonisation by persons with olfactory disorders.
Moreover, the device should not be operated in places directly exposed to sunlight, with high humidity, in
places where chemicals are used and in rooms with very small cubic capacity. For the correct operation of
the ozone sensors the device must be used in an environment with the following parameters:
Do not clean the room intended for ozone sterilisation with chlorine-based cleaning agents or other chlo-
rinated substances. This can cause erroneous indications of the ozone sensor in the device and expose the
user to danger.
The device is equipped with a pre-filter and an active carbon filter. In addition, there is one additional pre-
-filter in the package, which must be used interchangeably with the activated carbon filter when starting
the sterilisation mode.
Temperature: -20 to 50˚C
Pressure: 900 to 1100 hPa