24
9. Technical data
9.1 Device data
Manufacture in accordance with:
EU Directive 2014/35/EU
EU Directive 2014/30/EU
EU Directive 2014/68/EU
Applied standards:
EN 61000 -4
EN 61000 -6
DIN EN 12952-7
DIN EN 12953-6
EN 61010-1
EN 13445
Data exchange:
CAN-Bus in accordance with DIN ISO 11898,
CANopen protocol
Number of probes:
2 to 4
Total length of CAN-Bus:
max. 500m
Electrical conductivity of the liquid:
0.5 µS/cm ≤ æ ≤ 10,000 µS/cm (25°C)
Probes / water column:
Allowable pressure
PS
[bar]
32
200
Allowable temperature
TS
[°C]
239
367
Probe
Type
EL65
EL60
Item-no.
15-14864
15-13243
Insulator
PTFE
Ceramic
Connection thread
G ½″
Width across flats
AF27
Electrode spacing
At least 36 mm with offset arrangement
Material screw connection
Niro
Material electrode tip
Niro
Material water column
SA106 GrB (Standard); materials according to DIN or ASME
Process connection
according to DIN or ANSI; Flange or welding end