Mounting of the Aseptoflex adapter
Step 1
Grease thread and sealing cham-
fer of the sensor and of the
adapter with the greasing paste
supplied.
The greasing paste is food-grade
(USDA-H1 84-201).
Make sure that the O-ring (D) is
correctly positioned.
Step 2
Screw the sensor into the
adapter. Avoid mechanical influ-
ence on the sealing chamfers.
Step 3
Clamp sensor and adapter into a
clamping device (E). The sealing
chamfers (F) must not be dam-
aged. Tighten the sensor with a
spanner until you can feel the
end stop.
If you continue to turn, this
can have adverse effect on
the sealing.
24
I
E
F
36
D
Gr
ease