9.7 Setting of the preset meter / pulse value (ImPS)
7 setting ranges are available:
LED
Display
in steps of
Value
1
3
0 0� 0 1
���
9 9� 9 9
0�01 gal
0�01���99�99 gal
2
3
1 0 0� 0
���
9 9 9� 9
0�1 gal
100�0���999�9 gal
3
3
1 0 0 0
���
9 9 9 9
1 gal
1000���9999 gal
4
3 + 5
1 0� 0 0
���
9 9� 9 9
10 gal
10 000���99 990 gal
5
3 + 5
1 0 0� 0
���
9 9 9� 9
100 gal
100 000���999 900 gal
6
3 + 5
1 0 0 0
���
9 9 9 9
1000 gal
1 000 000���
9 999 000 gal
7
3 + 6
1 0� 0 0
���
9 9� 9 9
10 000 gal
10 000 000���
99 990 000 gal
Setting operation
:
► Set [OU1] to [ImP] (→ 8.3.2).
►
Press [Mode/Enter] until [ImPS] is displayed�
►
Press [Set] and keep it pressed�
> The current numerical value flashes for 5 s, then one of the 4 digits becomes
active (digit flashes, can be changed)�
►
Set the requested value as indicated in the following table�
►
First select the requested setting range� Then set the figure from the left digit to the
right digit�
►
Press [Mode/Enter] briefly when all 4 digits are set�
As soon as the first digit flashes there are 3 options:
►
Press [Set] once�
The flashing digit is increased� 9 is followed by 0 - 1 - 2, etc�
8 1� 2 3
1 x [Set] pressed 9 1� 2 3
1 x [Set] pressed 0 1� 2 3
1 x [Set] pressed 1 1� 2 3