SRX/SRC Series
8
Maintenance
4.3.3 Required
Equipment
Many combinations of standards, transfer stan-
dards, meters, and bridges may be used to calibrate
this instrument. The following are some possible
choices:
• Resistance Standards or Transfer Standards
for the required values with traceable cali-
brations, such as the following standards
available from IET Labs
• SR-102
100
Ω
•
SR-103 1 k
Ω
•
SR-104 10 k
Ω
• SRL
series
• Precision resistance measurement bridge
or multimeter, with a transfer accuracy
of ±l ppm. Options include:
• Guildline Model 9975
• Measurements
International
Model
6010C
• ESI model 242, 242A, 242C, or
242D
• A high-precision, high-stability digi-
tal multimeter (e.g. Fluke 8508A)
along with a set of resistance stan-
dards for ratio mode.
4.3.4 Calibration
Procedure
To calibrate an SRac unit, proceed as follows:
1. Set up the calibration equipment in the resis-
tance measurement mode.
2. Con
fi
rm the resistance of the unit.
3. Con
fi
rm that the resistance is consistent with
historical measurements.
4.4 Replaceable Parts List
Reference
IET Pt No
Description
1
BP-1000-RD
Binding Post, Red
2
BP-1000-BK
Binding Post, Black
3
BP-1000-GN
Binding Post, Green
Not Shown
SRac-*-Res
SRac resistor assembly
Replace * with nominal resistance value
Table 4-1: Replaceable Parts List
Figure 4-1: SRac Replaceable Parts
1