4.1  Initial Inspection and Setup

This instrument was carefully inspected before ship-

ment.  It should be in proper electrical and mechanical 

order upon receipt.
An  OPERATING  GUIDE  is  attached  to  the  case 

of  the  instrument  to  provide  ready  reference  to 

specifications.

4.2  Connection

Four insulated low-thermal-emf binding posts labeled 

CURRENT HI

CURRENT LO

SENSE HI

, and 

SENSE LO

 provide two current and two potential 

terminals, respectively, for 4-terminal measurement.  

2-terminal measurements may be made by shorting 

CURRENT HI

 to 

SENSE HI

, and 

CURRENT LO

 

to 

SENSE LO

, preferably with shorting links or other 

substantial means.  The four terminal connection is 

important of course for low resistances.

A fifth metal binding post labeled 

GND

 (Ground) is 

connected to the case and may be used as a guard or 

shield terminal.  It may also be connected using the 

shorting link to either terminal to implement a 2-ter-

minal as opposed to a 3-terminal measurement.

4.3  Electrical Considerations

In order to make proper use of the full performance 

capabilities of the HARS-LX unit, especially if low 

resistance or low resistance increments are important, 

care must be taken in connecting to the terminals of 

the decade box.

In particular, to keep contact resistance to a mini-

mum, 4-terminal connections should be employed, 

and the most substantial and secure connection to the 

binding posts should be made.  They accept banana 

plugs, telephone tips, spade lugs, alligator clips, and 

bare wire.  The largest or heaviest mating connection 

should be made, and, where applicable, the binding 

posts should be securely tightened.

These considerations may be relaxed whenever single 

milliohms not considered significant for the task be

-

ing performed.

4.3.1  Thermal emf Considerations

The highest quality low emf components are used in 

the HARS-LX series.  In particular, the terminals are 

made of gold plated tellurium copper, which exhibits 

low emf and low resistance.  There nevertheless may 

be some minute thermal emf generated at the user’s 

test leads where they contact the HARS-LX binding 

posts.  This will depend on the test lead material.  

Whenever this is critical, brass and iron materials 

should be avoided.

This emf will not manifest itself if an ac measurement 

instrument is employed.  It will also be eliminated if 

a meter with so called “True Ohm” capability is used.  

Otherwise it may appear as a false component of the 

dc resistance measurement, and can be the order of 

milliohms.

Summary of Contents for HARS-LX

Page 1: ...ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC HARS LX Laboratory Standard Decade Resistance Substituter User and Service Manual Copyright 2014 IET Labs Inc Visit www ietla...

Page 2: ...PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC...

Page 3: ...t no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the specifications shall void this warranty IET shall...

Page 4: ...USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme caution when working w...

Page 5: ...1 Thermal emf Considerations 5 4 4 Dial Setting 6 4 5 Power Considerations 6 4 6 Environmental Conditions 6 4 7 Switch Conditioning 6 4 8 Meter Shunt Applications 7 4 9 Kelvin Bridge Applications 7 C...

Page 6: ...tituter 1 Figure 2 1 Typical Operating Guide Affixed to Unit 3 Figure 4 1 Kelvin Bridge Connections 7 Table 5 1 Trimming Potentiometers 11 Figure 5 1 Typical Trimmer Board 11 Table 5 2 Replaceable Par...

Page 7: ...y feature solid silver alloy contacts and quadruple leaf silver alloy wipers which keep switch contact resis tance to under 1 m per decade and more impor tantly keep switch contact resistance reproduc...

Page 8: ...ermetically sealed low inductance Terminals Four 5 way gold plated tellutium copperbinding posts with low thermal emf and low resistance for four terminal Kelvin measurements plus one binding post con...

Page 9: ...00 1 211 0 M 4 100 100 HARS LX 4 1k 12 110 M 4 1 k 1 k HARS LX 4 10k 121 10 M 4 10 k none HARS LX 5 0 001 121 110 5 1 m HARS LX 5 0 01 1 111 10 k 5 10 m 10 m HARS LX 5 0 1 11 111 0 k 5 100 m 100 m HAR...

Page 10: ...100 short term average value Environmental Conditions Operating Temperature 0 C to 55 C Storage Temperature 40 C to 70 C Switch Type Multiple solid silver contacts dust tight diallyl phthalate body To...

Page 11: ...tandard temperatures near 23 C 3 3 Repackaging for Shipment If the instrument is to be returned to IET Labs contact the Service Department at the number or address shown on the front cover of this man...

Page 12: ...low resistance increments are important care must be taken in connecting to the terminals of the decade box In particular to keep contact resistance to a mini mum 4 terminal connections should be empl...

Page 13: ...0 Since the highest decade has an additional 11 po sition resistance values of over 1 2 times nominal maximum value can be obtained 4 5 Power Considerations To maintain the maximum possible accuracy a...

Page 14: ...resistance standard for all Kelvin bridge measurements When connected as shown in Figure 4 1 errors caused by lead and contact resis tances can be made negligible because they appear as part of the ge...

Page 15: ...igure 4 1 Figure 4 1 Rheostat Dial In order to eliminate contact resistance and thermal emf the HARS LX integrates the rheostat as shown in Figure 4 2 This way the wiper is in the low potential circui...

Page 16: ...from the side of the instrument and remove the housing 2 Remove any dust or debris using optical grade dry compressed air or a clean brush 3 Replace the housing and reattach the 4 hous ing screws The...

Page 17: ...he follow ing models HATS LR HATS Y SRL Series The 1 and 10 transfer standards are optional and are only used to take advan tage if desired of the adjustability of these two decades Precision resistan...

Page 18: ...trim or adjust any resistances proceed as follows 1 Stabilize HARS LX at laboratory tempera ture of 23 C for at least 8 hours 2 Remove the 4 housing screws from the sides of the instrument and remove...

Page 19: ...Designation All decades but 1 k Potentiometer Designation 1 k decade 1 T0 T1 2 T1 T2 3 T2 T3 4 T3 T4 5 T4 T5 6 T5 T6 7 T6 T7 8 T7 T8 9 T8 T9 10 T9 T10 11 T10 1 M steps only Table 5 1 Trimming Potenti...

Page 20: ...y Not Shown HARS 4000 LX 0 1 100 m step Decade Switch Assembly Not Shown HARS 4000 LX 1 1 step Decade Switch Assembly Not Shown HARS 4000 LX 10 10 step Decade Switch Assembly Not Shown HARS 4000 LX 10...

Page 21: ......

Page 22: ......

Reviews: