2 Connectors and Front Panel elements
All connectors and user interface elements are located on the front panel.
2.1 MAIN connector
The MAIN connector is used to connect this device to the user´s existing infrastructure.
Connections are:
Power supply to internal charger
Ethernet for communication
Trigger input
Sync input
The pinout is given in the table below.
Connector type in front panel: Lemo EGA.2B.318
Connector type for user cable: Lemo FGA.2B.318
Pin
Signal
Remark
1
Ethernet A-
2
Ethernet D-
3
Ethernet B-
4
Ethernet B+
5
Ethernet C+
6
7
+CHGSUP
Charger-Input 40-48 V, 1 A
8
+CHGSUP
as above
9
SYNC+
10
SYNC-
11
-CHGSUP
Device-GND and Charger-GND
12
-CHGSUP
as above
13
Ethernet A+
14
Ethernet D+
15
Ethernet C-
16
17
+T0
18
-T0
Case
The input signals SYNC and T0 are of type RS-485. They are pulled to their respective quiet
voltage levels by internal pullup resistors and therefore may be left open, if unused.