Operating Instruction:

           Level Gauge with Magnetic Flags 

   MAGTOP G5

 

 
SU0117b.DOC   009/04 

Subject to change without notice! 

 

 

 
 

Installation Procedure: 

 

MAGTOP G5 with Flanges: 

• 

Mount the top and bottom shut-off valves to the tank 
flanges using suitable sealing materials. Do not yet fully 
tighten the screws. 

• 

Insert the 11 x 12 x 5 silicon discs into the shut-off 
valves. 

• 

Push the sleeve nut, along with the silicon ring (which 
has been snapped into place inside the nut), over the 
standpipe (see figure. 2). 

• 

Insert the standpipe between the shut-off valves making 
sure that is square. 

• 

Check the length of the standpipe! The pipe must fit 
flush against the silicon discs. Loosen the flange screws 
if necessary to correct distance and squareness. 

• 

Retighten the flange screws. 

• 

Push the sleeve nut up to the shut-off valve and tighten 
(46 mm open-end wrench). 
Tighten the sleeve nut “hand tight” only. Do not crush 
the seal! 

• 

After the vessel has been filled, check all fittings for 
leaks! 

 

MAGTOP G5 with Threaded Connectors: 

• 

Mount the top and bottom shut-off valves to the 
threaded connectors at the tank using suitable sealing 
materials. 

• 

Insert the 11 x 12 x 5 silicon discs into the shut-off 
valves. 

• 

Push the sleeve nut, along with the silicon ring (which 
has been snapped into place inside the nut), over the 
standpipe (see figure. 2). 

• 

Insert the standpipe between the shut-off valves making 
sure that is square. 

• 

Check the length of the standpipe! The pipe must fit 
flush against the silicon discs. Correct if necessary. 

• 

Push the sleeve nut up to the shut-off valve and tighten 
(46 mm open-end wrench). 
Tighten the sleeve nut “hand tight” only. Do not crush 
the seal! 

• 

After the vessel has been filled, check all fittings for 
leaks! 

 
 

Maintenance: 
 

If used for its intended purpose, the MAGTOP G5 is 
maintenance-free. 
However, the magnetic float accumulates any magnetic 
particles which may be present in the medium over a 
period of time. In order to prevent operating errors, the float 
should be cleaned to remove magnetic particles at regular 
intervals. 

 

 

 
 
Figure 2: MAGTOP G5 Assembly with Flange 
Mounting (ME0440) 
 

 
 
 
 
 
 
 
 
 
 
 
 

 

 