Troubleshooting
102
5
Troubleshooting
5.1 Main sources of error
The following can cause error measurements or unwanted results:
▪
The patient’s arm moving during measurement
▪
Device being switched off (e.g. at night)
▪
Incorrect cuff size put on
▪
The cuff slipping while being worn
▪
Lack of successful manual measurement in the
doctor's practice
▪
Not taking the medication
▪
Incorrect log set
▪
Not fully charged, incorrectly charged or
outdated power packs used
▪
Kinking or knotting of the cuff tube
▪
External interference factors such as, for
example, physical activity or driving or using
public transport during measurement may
result in so-called motion artefacts or incorrect
measurements.
5.2 Transmission error
To avoid an error during data transmission, the blood pressure monitor includes a check of the data sent. If an
error has occurred, E004 will appear on the display.
5.3
Checklist
▪
Please go through the following checklist if you have encountered errors when handling the Mobil-O-
Graph
®
. Many errors have simple causes.
▪
Check that all cables are firmly connected and/or that the infrared interface is correctly aligned with the
device.
▪
Check if the blood pressure monitor, computer and printer are switched on (depending on the version you
have received).
▪
Check that the connection cable is connected to the correct serial interface (COM1 to COM4).
▪
Check that the power packs are sufficiently charged.
Summary of Contents for Mobil-O-Graph NG
Page 2: ...Revision B 17 10 2018 ...
Page 3: ...Gebrauchsanweisung DE Mobil O Graph PWA Mobil O Graph NG ...
Page 66: ...64 ...
Page 67: ...Operating manual EN Mobil O Graph PWA Mobil O Graph NG ...
Page 129: ...Manual de instrucciones ES Mobil O Graph PWA Mobil O Graph NG ...
Page 191: ...Mode d emploi FR Mobil O Graph PWA Mobil O Graph NG ...
Page 254: ...252 ...
Page 255: ...Istruzioni per l uso IT Mobil O Graph PWA Mobil O Graph NG ...
Page 320: ...318 ...
Page 321: ...Gebruiksaanwijzing NL Mobil O Graph PWA Mobil O Graph NG ...
Page 352: ...Productbeschrijving 350 Geschatte levensduur van de manchet 6 maanden ...
Page 384: ......
Page 385: ......