GeniSPOT Screen Calibration Kit
Page 31
5.2 Conformance Test
The conformance test compares the luminance read by the GeniSPOT to the luminance
that should be produced by the monitor at different luminance levels.
To perform the conformance test, follow these steps:
Step 1:
Click
C
ONFORMANCE
T
EST
in the main window. The conformance test starts
automatically.
Figure 5-2: Conformance Test
Step 2:
Place the GeniSPOT in the white block and click
OK
to continue.
Figure 5-3: Conformance Start
Step 3:
The GeniPASS software displays a white screen, then a black screen and then
progresses from black and through the grays to white. The process takes a few
minutes. No other software should be used while the conformance test is being
performed. The GeniSPOT must remain in the same position at the center of the
monitor throughout the test.
Figure 5-4: Conformance Test Progress
Summary of Contents for GeniSPOT
Page 1: ...GeniSPOT Screen Calibration Kit Page i...
Page 12: ...GeniSPOT Screen Calibration Kit Page xii List of Tables Table 2 1 Packing List Items 5...
Page 13: ...GeniSPOT Screen Calibration Kit Page 1 Chapter 1 1 Introduction...
Page 16: ...GeniSPOT Screen Calibration Kit Page 4 Chapter 2 2 Unpacking...
Page 18: ...GeniSPOT Screen Calibration Kit Page 6 Chapter 3 3 Installation...
Page 30: ...GeniSPOT Screen Calibration Kit Page 18 Chapter 4 4 Software Configuration...
Page 41: ...GeniSPOT Screen Calibration Kit Page 29 Chapter 5 5 Basic Functions...
Page 56: ...GeniSPOT Screen Calibration Kit Page 44 Chapter 6 6 Advanced Functions...
Page 61: ...GeniSPOT Screen Calibration Kit Page 49 Figure 6 9 Select Your Monitor...
Page 62: ...GeniSPOT Screen Calibration Kit Page 50 Appendix A A Compatibility...
Page 65: ...GeniSPOT Screen Calibration Kit Page 53 Appendix B B Hazardous Materials Disclosure...
Page 69: ...GeniSPOT Screen Calibration Kit Page 57 Index...