TOF-Controller Instructions for use - Rev D

2

Sensors Validation Test 

1/ Power up TOF-Controller using the power 

supply then select 

SENSOR.

2/ Connect Sensor to TOF-Controller then 

 

connect the electrodes clamps as shown 

above. 

3/ Position sensor as shown above   then 

press on TOF-Controller press 2s and release 

to start the test. Repeat the test using position 

.

Note: Sensor position for TOF-S2/B, TOF-PS/B, 

TOF-PS2/B is IDENTICAL as shown above. It is 

important to put the sensor on a hard surface 

to ensure perfect testing conditions.

4/ Position sensor as shown above   then 

press on TOF-Controller press 2s and release 

to start the test. Repeat the test using position 

.

Note: Sensor position for TOF-FS/B, TOF-ES/B, 

is IDENTICAL as above. It is important to 

 

immobilize the sensor on a hard surface to  

ensure perfect testing conditions.

OK

: Passing results

FAIL

: Please contact your local distributor or  

IDMED for further instructions.

Note: It is recommended to repeat the test to 

rule out any previous misuse.