Taking a measurement

Positioning of the device and detection of the pupil

 

Positioning the device on the patient’s face is an important step to obtain reliable 

measurements (see: “Installation and positioning of the AlgiScan” section).

 

If the patient is conscious, ask him/her to keep his/her head straight, open the filmed 

eye wide and look straight ahead without blinking. The other eye is closed.

 

If the patient is unconscious or not able to cooperate, the operator may need to  

raise and/or lower the patient’s eyelids so that the pupil measured by the AlgiScan appears 

completely unobstructed.

 

The patient’s pupil should be centered on the screen and totally in green color. 

Only the pupil will be colored in the image displayed by the AlgiScan. If this is not the case, 

 

reposition or position differently the AlgiScan.

 

It is important when measuring a very small pupil to position the AlgiScan so that the 

white spots (reflections of the infra-red diodes) do not deform the detected pupil contour. If 

this is the case reposition the AlgiScan or position it differently.

NO

White spots deform the pupil

YES

Correct positioning and

detection of the pupil

NO

Poor pupil detection

When the pupillometer is ready for a measure the following icon 

 will appear at the  

bottom right of the video image.

Note:

 when the pupil detection is not correct, the red icon will appear at the bottom right of the 

video image. The pupillometer does not detect the pupil in the video image.

Do not move the AlgiScan when taking measurements

Taking a measurement

Once the  device correctly positioned,  the AlgiScan displays the  green icon when the pupil 

is detected. It is necessary to ensure that the size of the pupil is stable before taking the  

measurement.

 

Start taking the measurement by pressing continuously on the main button until the 

countdown at the bottom right corner of the screen reaches 0. 

 

A beep will indicate the start and end of the measurement period. In “Flash” mode, at 

the end of the measurement, the user can record the patient’s eye used for the measurement 

(right or left). 

 

Bilateral Pupil Exam:

 If a measurement on the other eye is done within 2 minutes of 

the first one, a comparison chart of the bilateral pupil exam will be displayed.

Hold function

When a mode (“Flash”, “Tetanus”, “PPI” or “PRD”) is selected, it is possible to freeze the analysed 

image of the pupil by putting the AlgiScan on pause. To do this, press the video image for a 

long time, which freezes the measured pupil value as well as the filmed image. Press again on 

to deactivate the pause mode.

Summary of Contents for Algiscan

Page 1: ...User Manual AlgiScan 1 USER MANUAL AlgiScan Version 3 2 EN Last update 2021 04 19 Ref ALG IFU_EN USER MANUAL...

Page 2: ...User Manual AlgiScan 2...

Page 3: ...reen 12 General use 13 Fitting the positioning eyecup 13 Installation and positioning of the AlgiScan 13 Turning the device ON OFF 14 Patient Files Measurements identification 14 Settings 14 III Using...

Page 4: ...patient s file 29 IV Preventive maintenance cleaning and disinfection 30 Preventive maintenance 30 Battery and battery charging 30 Cleaning and Disinfection 31 V Appendix 1 32 Troubleshooting 32 VI A...

Page 5: ...also to the characteristics of its dynamic evolution minimum maximum latency and speed of evolution The size and reactivity of the pupil are considered as supplementary data which provide more precis...

Page 6: ...cessary to interpret with care values measured on patients suffering from neurological problems The AlgiScan conforms to the European directive covering medical devices and to the regulations in force...

Page 7: ...easurement Before and after the measurement the electric cable should be disconnected from the electrodes The AlgiScan must be connected to pacing electrodes or ECG electrodes supporting voltages up t...

Page 8: ...nt s face and head especially the eyes lips entire neck Do not position the electrodes on either side of the heart The use of the AlgiScan should not result in any significant pressure of its eyecup o...

Page 9: ...gesia and sedation of the patient as well as the site of stimulation We strongly advise against placing the electrodes on the patient s thorax because of the additional risks of cardiac fibrillation N...

Page 10: ...ds Not protected against liquids Refer to the user manual Type BF Applied Part Catalog reference Direct Current DC Prescription use only in the USA Curtis Straus Mark USA and Canada Manufacturing date...

Page 11: ...electrode symbol Correct electrode conduc tion level skin Yellow electrode symbol Moderate electrode conduc tion level skin Red electrode symbol Non functional electrical stimulation conduction probl...

Page 12: ...y screen in the Flash mode measurement mode used for measuring the pupillary light reflex Touch screen The AlgiScan is designed in such a way that all controls are accessible with a simple touch of a...

Page 13: ...by the user by selecting one of the available modes of the AlgiScan This type of measurement will then give a series of results with figures and a curve representing the variation in the size of the p...

Page 14: ...es numbered 1 to 20 If there is no barcode scanning the user will either choose one of those files or create additional files numbered 21 and above to save the measurement results Settings Setting the...

Page 15: ...enu follow the steps below Select settings menu by pressing on Then select data transfer by pressing on Using the arrows select the transfer mode PC HL7 Connect your receiver module on PC ref NL WDT o...

Page 16: ...upil detection When the pupillometer is ready for a measure the following icon will appear at the bottom right of the video image Note when the pupil detection is not correct the red icon will appear...

Page 17: ...is used to measure the pupil reactivity to a standardized electrical stimulation Before starting the user selects the intensity of the electrical stimulation which will be applied and makes sure it i...

Page 18: ...s strength is fixed at 320 Lux The overall duration of the test is 4 seconds Selection of light stimulation To select light stimulation and adjust its power Press the symbol on the screen representing...

Page 19: ...red in green on the screen Start recording by pressing the main button Press and hold the main button for the duration of the recording Recording stops automatically at 60s or when the user releases t...

Page 20: ...by pressing the navigation arrows until the desired value is displayed Confirm your selection by pressing the following symbol Note To exit the pop up window or cancel the intensity selection press th...

Page 21: ...tion is more than 13 during the 30 mA stimulation stage 40 6 The pupil s variation is more than 13 during the 40 mA stimulation stage 50 5 The pupil s variation is more than 13 during the 50 mA stimul...

Page 22: ...the end of the electrical stimu lation Note that the AlgiScan emits a first beep when the electrical stimulation starts Dur ing the stimulation the device emits a beep at each new level the tone of wh...

Page 23: ...ring Patient Electrodes conduction level and skin resistance The level of conduction of the electrodes is checked by the AlgiScan as soon as an electrical stimulation with a non zero current is select...

Page 24: ...displayed depend on the protocol used It is important to control the quality of the displayed curve and to verify that measurement values are not artefacts Thus the blue and black horizontal lines ind...

Page 25: ...ualitative scale of the photomotor reflex 5 levels 3 colours to determine commentary on pupil reactivity level The quantitative and qualitative scale of the photomotor reflex and resulting commentary...

Page 26: ...time It also displays the evolution of the variations in pupil size for the same file over time Normal zone strong reactivity Diameter 10 01 12h00 12h00 2 mm 4 6 8 Variation 10 20 30 40 History scale...

Page 27: ...olours The value taken into account for the qualitative quantitative scale of the Pupillary Reflex Dilation and for the display of the comment is the amplitude of the reflex value as a percentage They...

Page 28: ...itude of the reflex value as a percentage They are only indicative and must be corroborated with other physiological indicators commonly used Amplitude of the PRD reflex in PRD 5 5 PRD 12 12 PRD 20 PR...

Page 29: ...ds Barcode reading Create new patient s file Home page Note If a data transfer mode is active the following icon is displayed on the screen It allows send all the patient s data to the configured rece...

Page 30: ...ed to carry out repairs or maintenance operations Battery and battery charging Battery The AlgiScan contains a rechargeable Lithium Ion battery It is fitted with thermal protection and protection agai...

Page 31: ...Note Only qualified technicians or IDMED personnel are authorised to carry out repairs or maintenance operations on the battery Cleaning and Disinfection Attention Do not autoclave the AlgiScan or any...

Page 32: ...be cleaned with a lint free cloth and wiped carefully to avoid any stains or reflections The electrode wire used with the AlgiScan should not come into direct contact be submerged sprayed or filled wi...

Page 33: ...isposed of or recycled according to the regulations in force Specification and warranty Environment Storage and expedition The AlgiScan and its accessories should be stored or transported within the l...

Page 34: ...icker IEC 61000 3 3 Compliant The AlgiScan can be connected to the public mains network EMC Immunity Phenomenon Basic EMC standard Professional healthcare facility environment Immunity Test Levels Com...

Page 35: ...ce keep minimum separa tion from RF communication equip ment of 30cm RATED power frequency magnetic fields IEC 61000 4 8 30 A m 50 Hz or 60 Hz 30 A m 50 Hz or 60 Hz Mains power quality should be that...

Page 36: ...IAB Please check with your authorized distributor or with the manufacturer which products are available and approved in your country Light Stimulation Duration 1 second Intensity fixed at 320 Lux prec...

Page 37: ...MED references Medical accessories of the AlgiScan Reference Description STA W2 Wireless charging station for AlgiScan CAB STIM3 Electric stimulation cable for AlgiScan Other accessories Reference Des...

Page 38: ...User Manual AlgiScan 38...

Page 39: ...User Manual AlgiScan 39...

Page 40: ......

Reviews: