Operator Manual
Aspect 100-4T and ISR
L100-2843 Rev. E
30
Problem/Symptom
Possible Cause
Remedy
5. Table does not
elevate.
1. No Power.
2. Battery is depleted.
3. Emergency Stop
Pushbutton engaged.
4. Tabletop Motion
Control Panhandle
Switches disabled.
5. Panhandle Failure.
6. Control connection.
7. Actuator(s)
uncalibrated.
8. Actuator Failure.
1. Check electrical outlet.
2. Connect to AC Power
3. Reset Emergency Stop Pushbutton.
4. Reset Switch Lockout Pushbutton.
5. Test motion with Handset Control. If
elevate works correctly, replace
Panhandle Unit
6. Service Control connections.
7. Reinitialize Actuator Controller with
Handheld Control.
8. Replace Elevate Actuator.
6. Trendelenburg
motion does not
operate.
1. No Power.
2. Battery is depleted.
3. Emergency Stop
Pushbutton engaged.
4. Tabletop Motion
Control Panhandle
Switches disabled.
5. Panhandle Failure.
6. Control connection.
7. Actuator(s)
uncalibrated.
8. Actuator Failure.
1. Check electrical outlet.
2. Connect to AC Power.
3. Reset Emergency Stop Pushbutton.
4. Reset Switch Lockout Pushbutton.
5. Test motion with Handset Control. If
Trendelenburg works correctly, replace
Panhandle Unit.
6. Service Control connections.
7. Reinitialize Actuator Controller with
Handheld Control.
8. Replace Trendelenburg Actuator(s).
7. ISO-Centric Roll
does not operate.
(ISR only)
1. No Power.
2. Battery is depleted.
3. Emergency Stop
Pushbutton engaged.
4. Tabletop Motion
Control Panhandle
Switches disabled.
5. Panhandle Failure.
6. Control connection.
7. Actuator(s)
uncalibrated.
8. Actuator Failure.
1. Check electrical outlet.
2. Connect to AC Power.
3. Reset Emergency Stop Pushbutton.
4. Reset Switch Lockout Pushbutton.
5. Test motion with Handset Control. If
ISO-Centric Roll works correctly,
replace Panhandle.
6. Service Control connections.
7. Reinitialize Actuator Controller with
Handheld Control.
8. Replace ISO-Centric Roll Actuator.