Zeiss ZHE Cube Assembly Procedure 

 

 
 

 

 

11/12 

Step 

Action 

1.2.14

 

 

Orient the Exciter with the arrow pointing into the cube as shown below. 

 

1.2.15

 

 

Position the cube with the Exciter mounting area facing upward as shown below. 

Place the filter inside the Exciter mounting area with the arrow pointing in. Note that 

the filter shoulder side faces outward. 

 

Summary of Contents for Semrock FL CUBE EC P and C

Page 1: ......

Page 2: ...below 1 1 2 Remove both filter retainer rings from the cube assembly 1 1 2 1 The supplied retaining rings for the ZHE cube are installed shoulder side inward Remove the rings by gently grasping the sl...

Page 3: ...ly different sides to the retainer rings One is the flat Shoulder Side and one is the Open Side The Open Side ALWAYS faces into the cube assembly and is tightened against the installed filter The Shou...

Page 4: ...and the Dichroic is mounted diagonally at the center of the cube and shares the Emitter half of the assembly Filter Orientation The filters are oriented in the assembly as follows The Exciter is mount...

Page 5: ...the screws aside for reassembly later 1 2 3 Separate the cube halves to prepare for assembly of the Semrock filter set Note that there are two prongs used to locate and hinge the cube housing and a m...

Page 6: ...e that the clip locating tab is inserted into the notch at the side of the cube assembly Note Ensure the spring clip flanges face UPWARD 1 2 5 Reflective Coated Side The filter edge is concealed by th...

Page 7: ...ssembly to retain the Dichroic within the flanges of the spring clip Note The cube halves should fit together flush and without interference at either mounting surface Note DO NOT rest the cube assemb...

Page 8: ...2 8 Fasten the cube halves together using the two flat head screws removed previously Alternately tighten the screws until they are snug Once snug further tighten turn 1 2 9 Orient the Emitter with t...

Page 9: ...ul installation of the retaining ring to avoid cross threading Using your fingers carefully turn the ring while observing its position in the threads Keep the ring level across the threaded area while...

Page 10: ...THORLABS SPX602 Spanner Wrench The spanner wrench blades are shown below 1 2 13 Insert the spanner wrench blades into the slots of the Emitter retaining ring Tighten the ring snug against the Emitter...

Page 11: ...xciter with the arrow pointing into the cube as shown below 1 2 15 Position the cube with the Exciter mounting area facing upward as shown below Place the filter inside the Exciter mounting area with...

Page 12: ...s threading If the ring will not turn it is not threaded properly slowly turn the ring out and try again Do not force thread the ring as it will turn freely if oriented correctly Using the Spanner Wre...

Reviews: