High Voltage Warning (HI-V)
The meter LED remains RED and the beep lasts for 1 second and then goes silent
when voltages in excess of 30V are measured.
NOTE
: This feature does not work in the Ohm, capacitor, continuity or clamp
modes.
Data Hold Feature
Press the Hold button on the side of the meter to toggle in and out of the data
hold mode. “HOLD” appears in the upper left of the meter display when data hold
is active. Use the data hold feature to lock a measurement reading on the display.
Press the Hold button again to unlock the display and obtain a real-time reading.
Peak Feature (61-757 Only)
In PEAK mode, the meter displays the PEAK value of AC Volts or Current
measurement. PEAK function is enabled by Depressing the PEAK HOLD button
for > 2 seconds. To Exit PEAK Mode Depress the PEAK HOLD button again for > 2
seconds.
MAX/MIN Feature
The MAX/MIN records the maximum and minimum measured value. Press the
MAX/MIN button to activate this feature and to toggle between MAX, and MIN
readings. The unit will continually capture MAX and MIN values over time.
Depressing the MAX/MIN button for >2 sec. exits the MAX/MIN mode.
Relative Mode
Press “ ” button to enter the Relative mode. The “ ” symbol is displayed, and
the value on the display is stored as a reference value. In the Relative mode, the
value shown on the display is always the difference between the stored reference
value and the present reading. Press the “ ” button again to exit the Relative
mode.
Zero Feature
The “ZERO” button is used to zero out the display before measuring DC current.
Press the “ZERO” button to subtract out the non-zero number. Then, measure the
DC amps. Pressing the “ZERO” button again causes the “ZERO” to flash and the
original offset number to be displayed. Depress the “ZERO” button for >2 sec. to
exit this mode.