12
Resistance (Ohms)
Range
Resolution
MaxDisplay
Accuracy
400
Ω
0.1
Ω
400.0
±(0.7% +3)
4k
Ω
1
Ω
4.000
±(0.4% +3)
40k
Ω
10
Ω
40.00
400k
Ω
100
Ω
400.0
4M
Ω
1k
Ω
4.000
±(0.6%+3)
40M
Ω
10k
Ω
40.00
±(1.5% +5)
Open Circuit Voltage:
-1.3V approx.
To Measure Resistance:
1. Turn the power off to the circuit or device that is to be
measured and discharge all capacitors before attempting
a measurement.
2. Plug the black test lead into the COM port and the red test
lead into the V
Ω
Hz port.
3. Set the rotary switch to the
Ω
position.
4. For correct reading, ensure that the device being tested
contains no voltage.
5. Press the range button to select the proper range of the
meter.
6. Measure resistance.
11
Frequency with Hz Switch
Function
Range
Sensitivity
Accuracy
AC Voltage
400mV
40mV rms
0.01%
4V
0.2V rms
reading
40V
2V rms
+5 digits
400V
20V rms
750V
200V rms
AC Current
40mA
8mA rms
400mA
80mA rms
10A
8A rms
To Measure Frequency with the Hz Switch
This switch is used to quickly view frequency during a mea-
surement of AC voltage or current.
1. Push ~ Hz during an AC voltage or current measurement.
2. The LCD will display the frequency measurement of the cir-
cuit.
3. Push ~Hz again to return to the AC voltage or current mea-
surement.
Hz
True RMS
Hz
True RMS
Circuit Connection:
˜
˜