Battery Replacement  (refer to Figure 2) 

1. Disconnect the test leads from any circuit under test and turn off meter. 
2. Slide off the back case battery cover  
3. Remove battery from compartment and unsnap the battery connector. 
4. Install new 9V battery (NEDA #1604). An alkaline type is recommended.  
5. Install new battery into compartment using care not to pinch or bind battery leads. 
6. Slide battery cover along glide until cover snaps into bottom case.  
    Insure that cover aligned with snaps at bottom of the case. 
 

 

 

Replacing Fuse (refer to Figure 3) 

1.   Disconnect the test leads and turn the range switch to “OFF”.

 

2.   Remove the screw holding the bottom case cover, just above the battery cover. 
3.   The case bottom is secured to the case top by two internal snaps (at the LCD end).  
      lift up on the battery end until the case un-snaps.  
4.   Remove the defective fuse and replace with a 500mA, 250V Fast Acting fuse. 
      IDEAL #F-7 Type is recommended. 
5.   Replace bottom cover insuring that case is secured at the top by the internal snaps and  
      replace screw. 
 

 

 
Form number TM61603 

Rev 3 February 2002 

Summary of Contents for 61-603

Page 1: ...he following information Precautions and safety information Specifications Performance test procedure Calibration and calibration adjustment procedure Basic maintenance replacing the battery and fuses...

Page 2: ...tions 2 General Specification 2 Voltage Specifications 2 Current Specifications 2 Resistance Specifications 2 Diode Specifications 2 Battery Check Specifications 2 Performance Verification 3 Calibrati...

Page 3: ...as specified CAUTION These statements identify conditions or practices that could result in damage to the equipment or other property WARNING These statements identify conditions or practices that co...

Page 4: ...mA 250V fast acting fuse Dimensions 5 0 H X 2 8 W X 1 4 D 128mmH X 72mmW X 36mmD Weight Approximately 7 0 oz or 200g including battery RANGES and ACCURACY SPECIFICATION Function Setting Ranges Accurac...

Page 5: ...sures a minimum of 7 5V DC If the battery measures under 7 5V DC replace the battery before beginning the performance test 3 Input the values listed in Table 1 Table 1 Performance Verification Functio...

Page 6: ...un snaps 5 Using a calibrated meter ensure the battery measures a minimum of 7 5V DC If the battery measures under 7 5V DC replace the battery Calibration Procedure It is recommended that all IDEAL m...

Page 7: ...into bottom case Insure that cover aligned with snaps at bottom of the case Replacing Fuse refer to Figure 3 1 Disconnect the test leads and turn the range switch to OFF 2 Remove the screw holding th...

Reviews: