PERFORMANCE VERIFICAITONS

  

 
Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning 
correctly.  If the meter does not conform to any of the listed limits the calibration procedure must be 
performed. 
 

Performance Verification Preparation

 

 
1.   Turn on the Calibrator, allow calibrator to warm up. Temperature Stabilization 
      should be reached after 30 minutes.

 

2.   Remove battery cover and using a calibrated meter to ensure the battery measures a minimum of 7.5V 
      DC. If the battery measures under 7.5V DC, replace the battery before beginning the performance test. 
3.    Input the values listed in Table 1

 

 

Table 1  Performance Verification  

 

 

Function Setting 

 

 
Input 

 
Low Limit  

 
High Limit 

ACV 200 

190V AC@ 60Hz 

187.7 

192.3 

ACV 500 

400V AC@ 60Hz 

392 

408 

DCV 2  

1.900V DC 

1.888 

1.911 

DCV 20 

19.00V DC 

18.88 

19.11 

DCV 200 

190.0V DC 

188.8 

191.1 

DCV 900 

900V DC 

893 

906 

DCmA 2m 

1.900 mA DC 

1.879 

1.921 

DCmA 20m 

19.00 mA DC 

18.79 

19.21 

DCmA 200m 

190.0 mA DC 

187.9 

192.1 

DC A 10 

9.00 A DC 

8.80 

9.20 

 200 

100.0 98.7 

101.3 

 2K 

1.000K .992 

1.008 

 20K 

10.00K 9.92 

10.08 

 200K 

100.0K 99.2 

100.8 

 2M 

1.000M .988 

1.012 

 

 
Form number TM61603 

Rev 3 February 2002 

Summary of Contents for 61-603

Page 1: ...he following information Precautions and safety information Specifications Performance test procedure Calibration and calibration adjustment procedure Basic maintenance replacing the battery and fuses...

Page 2: ...tions 2 General Specification 2 Voltage Specifications 2 Current Specifications 2 Resistance Specifications 2 Diode Specifications 2 Battery Check Specifications 2 Performance Verification 3 Calibrati...

Page 3: ...as specified CAUTION These statements identify conditions or practices that could result in damage to the equipment or other property WARNING These statements identify conditions or practices that co...

Page 4: ...mA 250V fast acting fuse Dimensions 5 0 H X 2 8 W X 1 4 D 128mmH X 72mmW X 36mmD Weight Approximately 7 0 oz or 200g including battery RANGES and ACCURACY SPECIFICATION Function Setting Ranges Accurac...

Page 5: ...sures a minimum of 7 5V DC If the battery measures under 7 5V DC replace the battery before beginning the performance test 3 Input the values listed in Table 1 Table 1 Performance Verification Functio...

Page 6: ...un snaps 5 Using a calibrated meter ensure the battery measures a minimum of 7 5V DC If the battery measures under 7 5V DC replace the battery Calibration Procedure It is recommended that all IDEAL m...

Page 7: ...into bottom case Insure that cover aligned with snaps at bottom of the case Replacing Fuse refer to Figure 3 1 Disconnect the test leads and turn the range switch to OFF 2 Remove the screw holding th...

Reviews: