Page 18
Table 9: DC mA Test
Input
Reading
Step Current
1 10.000mA
9.940
to
10.060
2 100.00mA
99.40
to
100.60
Testing the Amp (A) Function
To verify the accuracy in the ampere (A) measurement function, do the
Following:
1. Connect the Calibrator to the
A
and
COM
inputs of the Meter.
2. Turn the rotary switch to
A
3. Apply the inputs for steps 1 - 6 in Table 10.
4. For each input, compare the readings on the Meter display to the readings
in Table 10.
5. If the display reading falls outside of the range shown in Table 10, the meter does not
meet specification.
Table 10. ACA Test
Input Reading
Step Current Frequency
1
1.0000A
60Hz
0.9870 to 1.0130
2
1.0000A
1KHz
0.9800 to 1.0200
3
1.0000A
3KHz
0.9720 to 1.0280
4 9.000A 60Hz 8.878
to
9.122
5 9.000A 1KHz 8.812
to
9.188
6 9.000A 3KHz 8.740
to
9.260
6. Press the blue button on the Meter to toggle to the
DCA
measurement function.
7. Apply the inputs for steps 1 - 2 in Table 11.
8. For each input, compare the reading on the Meter display to the readings for
your meter in Table 11.
9. If the display reading falls outside of the range shown in Table 11, the meter does not
meet specification.
Table 11. DCA Test
Input Reading
Step Current
1
1.0000A
0.9940 to 1.0060
2 9.000A
8.942
to
10.058
Testing the Frequency Function
To verify the accuracy of the Meter’s frequency function, do the following:
1. Connect the Calibrator to the
V
Ω
and
COM
inputs on the Meter.
Note
The accuracy of the Calibrator’s frequency function must be appropriate
for the specified accuracy of the Meter.
2. Set the rotary switch to
Hz
.
3. Set the Function Generator for the square wave voltage and frequency
for steps 1 - 6 of Table 12.