SignalTEK II
156810 Iss 1
User Guide
Page 21
Note that for the example of a Wiremap test shown above; if the
option ‘XOver Allowed’ had been checked
in the Wiremap SETUP options, the results would be displayed as shown in Fig 23 and Fig 24.
Fig 23
Fig 24
The wiremap tests may be run with no termination
– open, or with an Active Remote termination. When
connected, an image of an Active Remote will be shown on the display and its type identified. After a test
has been run, the length of the cable is displayed (range up to 100m (330ft)).
With an open termination the possible faults detected are:
Fig 25 Open circuit by pair
Fig 26 Short circuit by pin