Form Number TM61312-4
Rev 2 Aug 2007
Page 5
(3a) DC micro-amp and milli-amps (for 61-312 / 61-314)
Range Resolution
Accuracy
Input
Prodection
400.0µA 0.1
μ
A
4000µA 1
μ
A
40.00mA 10
μ
A
400.0mA 0.1mA
±(1.2% r 5 digits)*
500mA, 250V Fast
Blow Fuse
Overload Protection:
mA Input: 500mA, 250V Fast Blow fuse. (61-312 / 61-314)
(3b) DC Current (61-312 / 61-314)
Range Resolution
Accuracy
Input Protection
4.000A .001mA
10.00A* .01mA
±(1.5% r 5 digits)
10A, 250V
Fast Blow Fuse
Overload Protection:
A Input: 10A, 250V Fast Blow fuse. (61-312 / 61-314)
*
Caution
:
Do not make high current measurements on the 10A scale for longer that 15 seconds
This should be followed by a 15 minute cool down period. Exceeding 15 seconds may cause damag
to the meter and/or the test leads.
.
e
(4a) AC micro-amp and milli-amps Current (61-312 / 61-314)
Range Resolution
Accuracy
Input
Protection
400.0µA 0.1
μ
A
4000µA 1
μ
A
40.00mA 10
μ
A
400.0mA 0.1mA
±(1.5% r 5 digits)
*
1
40Hz
~
400Hz
500mA, 250V
Fast Blow Fuse
Overload Protection:
mA Input: 500mA, 250V Fast Blow fuse. (61-312 / 61-314)
(4b) AC Current (61-312 / 61-314)
Range Resolution
Accuracy
Input Protection
4.000A .001mA
10.00A* .01mA
±(2.0% r 5 digits)
*
1
40Hz
~
400Hz
10A, 250V
Fast Blow Fuse
AC
sin
*
1
t
fa
Overload Protection:
A Input: 10A (250V) Fast Blow fuse.
Conversion Type:
AC conversions are ac-coupled, true rms responding, calibrated to the
e wave input.
The specified accuracy is for sine wave at full scale and non-sine wave at half scale with cres
ctor up to 2.
*
Caution:
Do not make high current measurements on the 10A scale for longer that 15 seconds
This should be followed by a 15 minute cool down period. Exceeding 15 seconds may cause damag
to the meter and/or the test leads.
.
e