Form Number TM61312-4
Rev 2 Aug 2007
Page 17
Table 4c AC A Test:
Step
Range
Source Frequency
Reading
1
4A
3.000A
60Hz
2.935 to 3.065
2
4A
3.000A
400Hz
2.935 to 3.065
3
10A
10A
60Hz
9.75 to 10.25
4
10A
10A
400Hz
9.75 to 10.25
To
Testing the Resistance Function (for 61-312 / 61-314)
verify the accuracy of the resistance function, do the following:
1. Connect the calibrator to
V
Ω
Hz
and
COM
on the meter.
2. Turn the rotary switch to
Ω
.
3. Apply the inputs for steps 1 through 7 in Table 5.
4. Compare the meter display readings to the display readings in Table 5.
5. If the display reading falls outside of the range shown in Table 5, the meter does not
meet specification.
Table 5
Ω
Resistance Test:
Step
Range Source
Reading
1 400
Ω
* 4.0
Ω
3.6 to 4.4
2 400
Ω
* 300.0
Ω
296.6 to 303.4
3 4k
Ω
3.000K
Ω
2.966 to 3.034
4 40k
Ω
30.00K
Ω
29.66 to 30.34
5 400k
Ω
300.0K
Ω
296.6 to 303.4
6 4M
Ω
3.000M
Ω
2.960 to 3.040
7 40M
Ω
30.00M
Ω
29.50 to 30.50
*Lead resistance on the 400
Ω
range is not included in error.