Diana Peristaltic Pump, ICU Medical
Page 47
5.
Specifications, Maintenance and Troubleshooting
5.
All maintenance must be performed by an authorized ICU Medical representative except for routeine
cleaning and soecifically excluded activities.
ICU Medical recommends sending the unit to ICU Medical Service department once a year for
inspection and maintenance.
5.1.
Accuracy Specifications
Note:
Verify Diana Peristaltic Pump accuracy performance before use to ensure it meets
facility protocols.
Diana Peristaltic Pump accuracy with Water/Isopropyl Alcohol at a Slow-speed setting using
ICU Medical Luer-based components, 16G needle, 18G needle:
+/- 1% for volumes greater than or equal to 10 mL
+/- 2% for volumes greater than or equal to 5 mL and less than 10 mL
+/- 4% for volumes greater than or equal to 1 mL and less than 5 mL
+/- 10% for volumes greater than or equal to 0.5 mL and less than 1 mL
Diana Peristaltic Pump accuracy with Dextrose 10%/Lipids at a fluid-specific speed using ICU
Medical Lipids tubing set (Luer-based):
+/- 10% for volumes greater than or equal to 5 mL
Diana Peristaltic Pump can provide a sustained transfer rate (measured with Water) using
ICU Medical tubing set:
0.7 L/min with Luer-based sets
0.6 L/min with 16G needle
0.2 L/min with 18G needle
5.2.
Environmental Specification
Reference
Specification
Operating Temperature
+10°C to +40°C
Operating Humidity
15% to 60% (non-condensing)
Operating Altitude
0 to 2,000 meters above sea level
Storage Temperature
0°C to +60°C
Storage Humidity
15% to 85% (non-condensing)
Storage Altitude
0 to 2,000 meters above sea level
Transportation Temperature
0°C to +60°C
Transportation Humidity
15% to 85% (non-condensing)
Transportation Altitude
0 to 2000 meters above sea level
Physical Dimensions
5.7 in x 9.8 in x 8.6 in
Weight
11.2 pounds