Diana Peristaltic Pump, ICU Medical
Page 32
4.1.
Customization
The Customization tab allows the user to modify settings that are user-controllable.
4.1.1.
Custom Speed Settings
The user of Diana Peristaltic Pump can modify the speed settings of the pump in the Customization
tab. The unit allows the user to select up to 20 different speed settings. Unique named identifiers
can be created and matched to a specific speed setting (e.g., the highest flow rate is equal to pump
speed 20, while the slowest flow rate is equal to pump speed 1). The speed is measured in mL/min.
Default speed settings are show in the table below.
Note:
Precise pump speed is not user-configurable during compounding; only named identified
presets are available.
1.
In the Settings menu, select
“
Customization
”,
then select "Custom Speed Settings".
2.
A window will appear with the customization options for the Diana Peristaltic Pump peristaltic
pump stored in the system. In the next steps, user can change the name of the speed settings.
Speed settings can be activated or deactivated for user to see during fluid transfer.
Speed
mL/min
Slow (3)
165
Normal (9)
440
Fast (14)
700