6
SPECIAL POINTS
SPECIAL POINTS TO NOTE
Ensuring adequate ventilation
10.0 There must be good ventilation surrounding the
ICS PACIFIC
G400L Medical
fridge. Insufficient ventilation will cause the fridge to run outside the required
operating temperature range and can cause permanent and expensive damage to
both the compressor and vaccines.
For further information please contact the
ICS PACIFIC
service department.
Hi/Lo Temperature Light
11.0 When this indicator is illuminated the display is currently showing the High or Low
temperature reading. The reading is not indicative of the current temperature.
Refrigerating Light
12.0 This indicator may flash as part of its normal operation (It has nothing to do with the
temperature inside the fridge).
Defrost Light
13.0 This indicator shows the fridge is in defrost. The fridge will defrost every 6 hours.
This indicator will flash after defrost as part of its normal operation.
The UP/Light Button
14.0 The UP/Light button switches the internal light on and off.
The DOWN/Mute Button
15.0 The DOWN/Mute button is used to deactivate the alarm in the event it is activated.
The HI/LO Button
16.0 The Hi/Lo Button is used to check the maximum and minimum temperatures of the
fridge. It is also used to clear and reset these temperatures.
OTHER
The Alarm Function
The alarm will only operate if the cabinet’s temperature reaches above +8ºC or below
+2ºC for more than 30 minutes.
If the cabinet resumes normal operation, the alarm will automatically deactivate.
QCPP Accreditation
When performing Data Logging of temperature for QCPP Compliance ensure the
data logger is placed in the center of the fridge as per instructions provided with the
data logger from the Pharmacy Guild of Australia.
TEMPERATURE RANGE
+2
°
C to +8
°
C
@ 30
°
C Ambient temperature
+60% Humidity
ATTENTION PLEASE NOTE:
THIS REFRIGERATOR IS DESIGNED
TO OPERATE IN A ROOM TEMPERATURE
OF +5ºC TO +30ºC ONLY.