background image

 

SPECIAL POINTS 

 

SPECIAL POINTS TO NOTE 

Ensuring adequate ventilation

 

10.0  There must be good ventilation surrounding the 

ICS PACIFIC 

G400L Medical 

fridge. Insufficient ventilation will cause the fridge to run outside the required 
operating temperature range and can cause permanent and expensive damage to 
both the compressor and vaccines. 

For further information please contact the 

ICS PACIFIC

 service department. 

 

Hi/Lo Temperature Light

 

11.0  When this indicator is illuminated the display is currently showing the High or Low 

temperature reading. The reading is not indicative of the current temperature. 

 

Refrigerating Light

 

12.0  This indicator may flash as part of its normal operation (It has nothing to do with the 

temperature inside the fridge). 

 

Defrost Light

 

13.0  This indicator shows the fridge is in defrost. The fridge will defrost every 6 hours. 

This indicator will flash after defrost as part of its normal operation. 

 

The UP/Light Button

 

14.0  The UP/Light button switches the internal light on and off.  
 

The DOWN/Mute Button

 

15.0  The DOWN/Mute button is used to deactivate the alarm in the event it is activated.  
 

The HI/LO Button

 

16.0  The Hi/Lo Button is used to check the maximum and minimum temperatures of the 

fridge. It is also used to clear and reset these temperatures. 

OTHER 

 

The Alarm Function 

 

The alarm will only operate if the cabinet’s temperature reaches above +8ºC or below 
+2ºC for more than 30 minutes.  

If the cabinet resumes normal operation, the alarm will automatically deactivate.  

 

 

QCPP Accreditation 

When performing Data Logging of temperature for QCPP Compliance ensure the 
data logger is placed in the center of the fridge as per instructions provided with the 
data logger from the Pharmacy Guild of Australia. 

 

TEMPERATURE RANGE 

 
+2

°

C to +8

°

@ 30

°

C Ambient temperature 

+60% Humidity

 

ATTENTION PLEASE NOTE:  

THIS REFRIGERATOR IS DESIGNED 

TO OPERATE IN A ROOM TEMPERATURE 

OF +5ºC TO +30ºC ONLY. 

 

Summary of Contents for G400L

Page 1: ...Manual 31 Healey Road Dandenong South Victoria 3175 Australia P O Box 4140 Dandenong South BC Victoria 3164 Australia E mail info icspacific com au Web www icspacific com au Phone 61 03 9706 5115 Facsimilie 61 03 9706 6556 Toll Free Sales 1800 664 664 Service 1800 630 360 ...

Page 2: ...__ These numbers can be found on the back of the cabinet TABLE OF CONTENTS Introduction 2 Safety Instructions 3 Safety Instructions 3 General Instructions 4 Setting Up The Cabinet 4 Stocking The Cabinet 4 General Operating Instructions 4 Cleaning Instructions 4 Electronic Digital Controller 5 Understanding The Electronic Digital Controller 5 To Operate The Light 5 Reading And Resetting The High Lo...

Page 3: ...he cabinet off at the power point first and remove the 3 pin plug carefully from the power point Always disconnect the 3 pin plug of the cabinet from the power point before cleaning to avoid danger of electrocution Do not store volatile or flammable substances inside of the cabinet Do not place any pressurised containers into the cabinet There is risk of explosion or fire Do not climb on or inside...

Page 4: ...helf When stocking the shelves ensure that stock is not overhanging the sides of the shelves to ensure correct air circulation Stock must not touch the front glass internal side panels When using containers such as boxes or trays on the shelves ensure there is adequate ventilation in the bottom of the containers and around the containers so that airflow is not restricted If in any doubt whatsoever...

Page 5: ...values press the HI LO Button and hold it in for 6 to 7 seconds The reset will be signalled by one long beep 4 Every time the high and low temperatures are checked it is important to clear and reset the displayed values Alarm function 1 The alarm will activate if the temperature inside the fridge goes below 2ºC or rises above 8ºC for longer than 30 minutes The alarm will activate and the display w...

Page 6: ...rs This indicator will flash after defrost as part of its normal operation The UP Light Button 14 0 The UP Light button switches the internal light on and off The DOWN Mute Button 15 0 The DOWN Mute button is used to deactivate the alarm in the event it is activated The HI LO Button 16 0 The Hi Lo Button is used to check the maximum and minimum temperatures of the fridge It is also used to clear a...

Page 7: ...within the evaporator Ensure that the drain tube has not dislodged from the drip tray at the rear of the cabinet If there is condensation on the cabinet The ambient room temperature must not be above 30 C The relative humidity must not be above 60 The door must remain securely closed when not in use If the cabinet is running warm inside The ambient room temperatures must not be above 30 C The rela...