4
GENERAL INSTRUCTIONS
SETTING UP THE CABINET
•
Before using check to ensure that there is no damage to the cabinet.
•
Remove all packaging, including any protective film.
•
Select a location without direct exposure to sunlight.
•
Select a location away from all heat and moisture generating equipment or sources.
•
The cabinet must be installed on a flat and level surface to ensure correct operation.
•
Installation must not be in a humid and / or dusty area.
•
Ensure there is ample clearance behind the cabinet to allow correct airflow to the cabinet.
•
A minimum of 10cm clearance must be between the cabinet and the sides / rear wall and a
minimum of 2cm on the top of the cabinet to allow adequate air circulation. Insufficient
ventilation will cause the fridge to run outside the required operating temperature range and
can cause permanent and expensive damage to both the compressor and vaccines.
•
It is essential that the shelves have been fitted correctly during use.
•
Ensure that each shelf is inserted with the higher rear guard at the back.
•
Incorrectly fitted shelves disrupt the air flow in the fridge causing erratic temperature and
QCPP Compliance failure.
•
Connect the power cord into a grounded, 3 pin, 10AMP power point that has appropriate power surge protection.
•
Ensure that your cabinet is connected to its own dedicated power point.
Do not use an extension cord, double adaptor or power board.
STOCKING THE CABINET
•
When stocking the top shelf, ensure that there is adequate (at least 10cm) ventilation beneath the fan
cover, which is located above the top shelf.
•
When stocking the shelves ensure that stock is
not
overhanging the sides of the shelves to ensure
correct air circulation. Stock
must not touch
the front glass / internal side panels.
•
When using containers such as boxes or trays on the shelves ensure there is adequate ventilation in
the bottom of the containers and around the containers so that airflow is not restricted. If in any doubt
whatsoever contact the Vaccine Supplier.
GENERAL OPERATING INSTRUCTIONS
•
Plug the cabinet into the wall and turn on at the power point.
•
After turning the cabinet on, allow the cabinet temperature to stabilize for 1~2 hours before stocking
with vaccines / product.
•
Manually defrost the cabinet if and when required.
•
The
ICS PACIFIC
G400L Medical fridge is pre-set to operate within a temperature range of +2Cº
and +8ºC in accordance with QCPP Accreditation requirements.
•
For controller operating instructions please see the “Electronic Digital Controller” on Page 5.
CLEANING INSTRUCTIONS
•
Turn the cabinet off at the power point and disconnect the 3-pin plug from the wall.
•
Before cleaning the cabinet remove all products from within the cabinet.
•
When cleaning the cabinet interior / exterior it is recommended that you use a solvent of warm water
and mild detergent or a food grade cleaning product. Never use any abrasive products.
•
The
ICS PACIFIC
G400L Medical refrigerator should be cleaned regularly. Aim to clean behind the
refrigerator at least every 6 months to prevent a dust / dirt build up.
•
Ensure that the back of the refrigerator is kept clean of dust and dirt. Use ONLY a brush or vacuum
cleaner to remove any dust or dirt build up that may occur.
Do not use water
.
•
Wipe the inside of the refrigerator with a damp cloth.
DO NOT USE
anything abrasive or corrosive.