6 Set up with the display and adjustment module
WIKA Operating Instructions - Process pressure transmitter CPT-2x
Diagnosis:
Information, e.g. on instrument status, pointer, measure-
ment reliability, simulation
Additional adjustments:
PIN, date/time, reset, copy function
Info:
Instrument name, hardware and software version, date of manu-
facture, sensor features
Note:
For optimum adjustment of the measuring point, the individual sub-
menu items in the main menu item "
Setup
" should be selected one
after the other and provided with the correct parameters. If possible,
go through the items in the given sequence.
The submenu points are described below.
6.5.1 Setup
In the menu item "
Sensor TAG
" you edit a twelve-digit measurement
loop designation.
You can enter an unambiguous designation for the sensor, e.g. the
measurement loop name or the tank or product designation. In digital
systems and in the documentation of larger plants, a singular desig-
nation must be entered for exact identification of individual measuring
points.
The available digits include:
•
Letters from A … Z
•
Numbers from 0 … 9
•
Special char, -, /, -
In this menu item you activate/deactivate the slave sensor for elec-
tronic differential pressure and select the application.
CPT-2x can be used for process pressure and level measurement.
The setting in the delivery status is process pressure measurement.
The mode can be changed in this adjustment menu.
If you have connected
no
slave sensor, you confirm this with "
Deac-
tivate
".
Depending on the selected application, different subchapters in the
following adjustment steps are important. There you can find the
individual adjustment steps.
Enter the requested parameters via the appropriate keys, save your
settings with
[OK]
and jump to the next menu item with the
[ESC]
and
the
[->]
key.
Measurement loop name
Application