background image

Copyright ©  2016 ICP DAS Co., Ltd. All Rights Reserved              E-mail: [email protected] 

 

 

1  

Mounting the Hardware

 

The  PMD  can  be  mounted  on  a  panel  of  maximum  thickness  5  mm. 

Adequate access space can be available at the rear of the instrument panel 

for wiring and servicing purposes.

 

 

 

To ensure proper ventilation for your PMD,   

leave a minimum of 50 mm space between the top 

and bottom edges of the PMD and the enclosure 

panels. 

 

 

 

 

 

 

 

 

 

 

1) Prepare the panel and cut the hole to the 

specified size. The table below shows the 

dimension of the panel cut-out of PMD. 

 

   

Models 

Depth 

PMD-220x 

44 mm 

PMD-420x 

54 mm 

Models 

Panel Cut-Out (Width x Height) 

PMD-220x 

197 mm x 133 mm, + 1 mm 

PMD-420x 

277 mm x 215 mm, + 1 mm 

Summary of Contents for PMD

Page 1: ...en download html ftp ftp icpdas com pub cd pmd_cd PMD Official Website PMD 2201 http pmms icpdas com en PMD_2201 html PMD 4201 http pmms icpdas com en PMD_4201 html PMD 2206 http pmms icpdas com en PM...

Page 2: ...for wiring and servicing purposes To ensure proper ventilation for your PMD leave a minimum of 50 mm space between the top and bottom edges of the PMD and the enclosure panels 1 Prepare the panel and...

Page 3: ...s Reserved E mail service icpdas com 3 2 Attach the PMD to the cut out hole 3 Insert the panel mounting clips into the upper and lower ventilation holes 4 Screw the panel mounting clips to the panel R...

Page 4: ...at the first time be sure that the rotary switch is placed in the 0 position 3 Connect to PC Network and Power The PMD is equipped with an RJ 45 Ethernet port for connection to an Ethernet hub switch...

Page 5: ...Gateway 192 168 0 1 DNS Server 8 8 8 8 Default Google DNS Server The user can modify the Network setting of PMD by two interfaces local side display interface and remote side Web page interface The N...

Page 6: ...68 255 10 Subnet mask 255 255 0 0 Gateway address 192 168 0 1 Step2 Make sure the PC and PMD LAN1 is connecting through Ethernet Step3 Open a PC side browser and type in the URL address http 192 168 2...

Page 7: ...a browser and type in the URL address of the PMD Open a browser and type in the URL address The new IP setting of the PMD in the address bar Make sure the IP address is accurate Step 2 Get on the PMD...

Reviews: