A/D Trigger Mode
1.
Software A/D Data Acquisition Mode
The A/D acquisition parameters are configured via a command from
the Host PC. The continuous A/D acquisition or the acquisition of N
data samples begins after the command is triggered.
2.
External Clock Signal Synchronization A/D Conversion Data
Acquisition mode
The speed of the A/D acquisition and the amount of data acquired are
controlled by external electrical signals. A falling edge for each output
waveform triggers an AD conversion
3.
External Digital Signal Event Trigger Mode
i. Pre-Trigger (acquisition of N data samples)
The A/D data is continually collected and is temporarily stored in the
memory on the PET-7H16M until the trigger signal is received. Once
the trigger signal is received, the collected N data samples are then
transferred to the Host PC.