![background image](http://html1.mh-extra.com/html/icp-das-usa/pcie-lm4/pcie-lm4_user-manual_4476325044.webp)
High-speed Multifunction Boards
User Manual, Ver. 1.0, Jan 2020, PMH-032-10 Page 44
Step 6:
Calibrate the Analog Input Channel 4 to 9.9V
(1)
Connect 9.9 V voltage source to PCIe-LM4.CON1.AI4+ (Pin14)
(2)
Connect GND source to PCIe-LM4.CON1.A14- (Pin48)
(3)
Click CH4 Rate Group 0
“Read”
button to get hexadecimal value
(4)
Click CH4 Rate Group 1
“Read”
button to get hexadecimal value
(5)
Click CH4 Rate Group 3
“Read”
button to get hexadecimal value