pH Buffer Reading vs Temperature
Temp(°C)
pH 4.01
pH 6.86
pH 9.18
pH 4.00
pH 7.00
pH 10.01
0
4.01
6.98
9.47
4.01
7.12
10.32
5
4.01
6.95
9.38
4.00
7.09
10.25
10
4.00
6.92
9.32
4.00
7.06
10.18
15
4.00
6.90
9.27
4.00
7.04
10.12
20
4.00
6.88
9.22
4.00
7.02
10.06
25
4.01
6.86
9.18
4.00
7.00
10.01
30
4.01
6.85
9.14
4.01
6.99
9.97
35
4.02
6.84
9.10
4.02
6.98
9.93
40
4.03
6.84
9.07
4.03
6.97
9.89
45
4.04
6.83
9.04
4.04
6.97
9.86
50
4.06
6.83
9.01
4.06
6.97
9.83
55
4.08
6.83
8.99
4.07
6.97
9.91
60
4.10
6.84
8.96
4.09
6.98
9.97
70
4.12
6.85
8.92
4.12
6.99
9.76
80
4.16
6.86
8.89
4.16
7.00
9.74
90
4.01
6.88
8.85
4.20
7.02
9.73
ORP Calibration: Press the
[MENU] button to enter the calibration mode, select ORP standard liquid calibration, and input the known
standard liquid value (Default 235mV).
Press the
[MENU] button to return to the measurement screen and press the
[CAL] button to enter the standard solutioncalibration mode. If there is no need to modify the value, this setting is omitted.
Press the
[CAL] button to enter the calibration mode. After entering the calibration mode, the instrument will be displayed as the upper
right image, and the corresponding mV value will be displayed on the left side of the screen.
After the calibration is completed, the slope will be displayed on the right side of the screen. If the instruemnt has been calibrated, press the
[CAL] button to inspect the calibration state, and then press the
[ENT] button to enter the re-calibration.
Field Calibration
Summary of Contents for proCon D500 Series
Page 1: ...D500 D520 Series pH ORP Digital Analyzer Controller Operating Manual...
Page 7: ...D500 D520 Series Operating Manual 7 Slim Line Wall Mount Bracket Typical Application...
Page 15: ...D500 D520 Series Operating Manual 15...
Page 16: ...D500 D520 Series Operating Manual 16 Measurement Settings...
Page 17: ...D500 D520 Series Operating Manual 17 SimplCal Calibration...
Page 18: ...D500 D520 Series Operating Manual 18 Field Calibration...
Page 19: ...D500 D520 Series Operating Manual 19 Calibration Adjusment...
Page 20: ...D500 D520 Series Operating Manual 20 Alarm Relay Settings...
Page 21: ...D500 D520 Series Operating Manual 21 Output Settings...