background image

FMX 320/FMX 640  SERIES IR CAMERA USER MANUAL

6

FMX 320 S

FMX 640 S

Detector Array

UFPA (VOx)

Pixel Pitch

17 μm

FOV

Lens dependent

Measurement Distance

Lens dependent

Pixel Resolution

336 x 256

640 x 512

Spectral Band

7 μm to 14 μm

Thermal Sensitivity (NETD)

< (30 mK) 0.03 °C at 30 °C (86°F)

Frame Rate

7.5 Hz

Dynamic Range

14-bit

Temperature Range

25 °C to 45 °C (77 °F to 113 °F)

Operation Range

0 °C to 50 °C (32 °F to 122 °F)

Storage Range

-40 °C to 80 °C (-40 °F to 176 °F)

Humidity

 

5% to 95% non-condensing

Accuracy

± 0.2 °C (0.36 °F)

Pixel Operability

> 99 %

Dimensions (without lens)

157 mm x 70 mm x 70 mm (L x W x H ± .5 mm)

(6.18” x 2.76” x 2.76” (L x W x H ± 0.02”))

Power

5 V DC 2 A, < 10 W

Weight (without lens)

< 520 g (1.15 lbs)

Interface

RJ45/Et PoE

Video

Raw

Emissivity Correction

0.01 to 1.0

IP Rating

IP 54

Shutter

Built-in shutter

SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE

Summary of Contents for FMX 320 Series

Page 1: ...Export Control THIS DEVICE IS INTENDED FOR ADJUNCTIVE USE WITH OTHER CLINICAL DIAGNOSTIC PROCEDURES TO MEASURE HUMAN BODY TEMPERATURE VIA NON CONTACT SKIN MEASUREMENTS VISUALIZED FROM THE HUMAN FACE NOT MEANT FOR STANDALONE CLINICAL DIAGNOSTIC PROCEDURES OR TO TREAT OR DIAGNOSE PATIENTS For more information contact us 409 861 0788 sales infraredcameras com www infraredcameras com Medical Medical E...

Page 2: ...CURATE BUT ARE PRESENTED WITHOUT WARRANTY OF ANY KIND NOTWITHSTANDING ANY OTHER WARRANTY HEREIN ALL DOCUMENT FILES AND SOFTWARE ARE PROVIDED AS IS WITH ALL FAULTS ICI DISCLAIMS ALL WARRANTIES EXPRESSED OR IMPLIED INCLUDING WITHOUT LIMITATION THOSE OF MERCHANTABILITY FITNESS FOR A PARTICULAR PURPOSE AND NONINFRINGEMENT OR ARISING FROM A COURSE OF DEALING USAGE OR TRADE PRACTICE THE SOFTWARE LICENSE...

Page 3: ...Safety Information 3 Intended Use 4 Technical Specifications 5 Structure 6 Package Contents 7 Installation Instructions 8 Computer Setup 9 Cleaning and Maintenance 10 Site and System Validation Checklist 11 Troubleshooting 12 About ICI Contents 4 4 4 5 7 8 9 16 18 19 22 24 ...

Page 4: ... to prevent dangerous actions such as the device being hit by external force or falling from heights During transportation and storage the original packaging box must be used Prior to start of the device make sure that the power supply is properly connected If the power supply is connected incorrectly the device may be damaged Do not place any objects on the power cord and do not place the device ...

Page 5: ...amic Range 14 bit Temperature Range 25 C to 45 C 77 F to 113 F Operation Range 0 C to 50 C 32 F to 122 F Storage Range 40 C to 80 C 40 F to 176 F Humidity 5 to 95 non condensing Accuracy 0 2 C 0 36 F Pixel Operability 99 Dimensions without lens 157 mm x 70 mm x 70 mm L x W x H 5 mm 6 18 x 2 76 x 2 76 L x W x H 0 02 Power 5 V DC 2 A 10 W Weight without lens 520 g 1 15 lbs Interface RJ45 Ethernet Po...

Page 6: ...t Temperature Range 25 C to 45 C 77 F to 113 F Operation Range 0 C to 50 C 32 F to 122 F Storage Range 40 C to 80 C 40 F to 176 F Humidity 5 to 95 non condensing Accuracy 0 2 C 0 36 F Pixel Operability 99 Dimensions without lens 157 mm x 70 mm x 70 mm L x W x H 5 mm 6 18 x 2 76 x 2 76 L x W x H 0 02 Power 5 V DC 2 A 10 W Weight without lens 520 g 1 15 lbs Interface RJ45 Ethernet PoE Video Raw Emis...

Page 7: ...FMX 320 FMX 640 SERIES IR CAMERA USER MANUAL 7 5 Structure 5 1 Appearance 5 2 Definitions of Housing Interface 1 Mirco USB interface 2 RJ 45 Internet access 1 2 ...

Page 8: ... ENSURE ALL SYSTEM EQUIPMENT AND COMPONENT ITEMS ARE PRESENT BEFORE BEGINNING INSTALLATION FMX Series IR Camera AC Power Cord AC Power Adapter Laptop Temperature Reference AC Power Cord Software USB Drive Option 1 Combo Cable Option 2 Power Adapter and Ethernet Cable ...

Page 9: ...OORWAYS AND AIR CONDITIONING VENTILATION SYSTEMS THE SELECTED AREA SHOULD HAVE A STABLE AMBIENT TEMPERATURE BETWEEN 20 C AND 24 C 68 F TO 75 2 F AND RELATIVE HUMIDITY RANGE FROM 10 TO 50 7 1 Option One Setup A B Plug AC power cord into AC power adapter Plug power adapter into laptop C Plug power cord into a 110 120V electrical outlet ...

Page 10: ...320 FMX 640 SERIES IR CAMERA USER MANUAL 10 Mount the Temperature Reference to a tripod using the 20 mount D E Plug power cord into Temperature Reference F Plug power cord into a 110 120V electrical outlet ...

Page 11: ...S TO BE IMAGED TO ENSURE THE EQUIPMENT WILL NOT BE MOVED OR KNOCKED DOWN USING A DIVIDING BARRIER WILL HELP KEEP TRIPODS SEPARATE FROM THE PATH G H Plug Mirco USB end of the combo cable into the 5V DC port on the back of the FMX Series IR Camera Mount the FMX Series IR Camera to a tripod using the 20 mount ...

Page 12: ...ower on devices ALLOW AT LEAST 45 MINUTES FOR THE TEMPERATURE REFERENCE SOURCE TO WARM UP THIS WILL PROVIDE THE MOST ACCURATE SKIN TEMPERATURE MEASUREMENTS SOFTWARE ARRIVES INSTALLED ON COMPUTERS AND TABLETS PURCHASED FROM ICI AS COMPLETE SYSTEMS AND IT WILL LAUNCH AUTOMATICALLY ON STARTUP AFTER THE SOFTWARE LAUNCHES THE USER SHOULD ENSURE THE TARGET ZONE IS IN VIEW OF THE CAMERA AND THE TEMPERATU...

Page 13: ...et cable into the FMX Series IR Camera Ethernet port Plug the other end of the Ethernet cable into the Ethernet port of the laptop computer K Plug power cord into a 110 120V electrical outlet SOME FMX SERIES IR CAMERA PACKAGES WILL COME WITH A POWER ADAPTER AND ETHERNET CABLE INSTEAD OF A COMBO CORD FOLLOW STEPS A G FOR OPTION ONE SETUP BEFORE CONTINUING WITH THE FOLLOWING STEPS ...

Page 14: ...on devices FMX 320 FMX 640 SERIES IR CAMERA SETUP GUIDE 2105 W Cardinal Dr Beaumont TX 77705 Phone 409 861 0788 Toll Free 866 861 0788 infraredcamerasinc com sales infraredcamerasinc com We Are IR 7 3 Full Assembly Diagram GRAPHIC FOR ILLUSTRATION PURPOSES ONLY ...

Page 15: ...emoval of glasses headwear and masks Read the current government guidance regarding the use of telethermographic systems for skin temperature measurements Additional information can be found by reading IEC 80601 2 59 2017 Medical electrical equipment Part 2 59 Particular requirements for the basic safety and essential performance of screening thermographs for human febrile temperature screening an...

Page 16: ...eed to set a gateway as this is a direct connection from the computer to the cameras with no router Be sure the static IP address is set by doing the following 1 Click on Windows icon 2 Click on Settings 3 Click on Network and Internet 4 Click on Ethernet 5 Click on Change Adapter Settings 6 Right click on Ethernet and choose properties 7 Double click on Internet Protocol Version 4 TCP IPV4 Proper...

Page 17: ...ting the system s computer to the Internet or your secure internal network be sure to have your IT department install your organization s anti virus software anti malware and if applicable system access software and apply security updates as necessary If connecting the system s computer to the Internet be sure to locate the computer behind a firewall Install security updates to the computer s oper...

Page 18: ... the middle to the edge paying attention to not crack the lens or use too much liquid If the lens is still not clean replace the cloth and repeat the wiping process 9 Cleaning and Maintenance 9 2 Disinfecting the Camera Surface Do not use corrosive cleaning solutions on the optical glass components It is recommended to disinfect the camera surface regularly with a non corrosive sanitizing product ...

Page 19: ... airflow HVAC vents intakes Any radiant energy from electrical sources Direct or indirect lighting on individual being screened Room temperature is 20 C 24 C 68 F 75 F Relative humidity is within 10 50 10 2 The Screening Background Placement is parallel and perpendicular to screening camera At a distance of 12 feet Consists of a non IR reflective homogeneous background Minimum size is 10 x 8 White...

Page 20: ... of footprints The distance is unique to each screening system Ensure the line of sight from the Temperature Reference Source to the camera is never blocked by an individual entering or leaving the point of measurement The secondary screening area should be setup near the screening area 10 4 Typical traffic flow illustration 10 5 IR Camera Placement Level on tripod or other suitable support Perpen...

Page 21: ...e On tripod or other stable stand in front of background and positioned for visibility within the frame of the image Temperature is set at 35 C 95 F or 37 C 98 6 F Warm up time is 15 minutes 10 7 Software Set up the software according to the instructions Documentation can be found online on the IR Flash STM web page under the Downloads section or at this address https infraredcameras com ir flash ...

Page 22: ...IR Flash STM software 11 2 Camera out of focus For FMX 320 and 640 cameras Adjust focus ring Close and reopen software Try newer or different version of IR Flash STM software Restart computer 11 3 Temperature readings are incorrect or facial recognition is suboptimal Close and reopen software Check that software settings are correct Fixed temperature of 35 C or 37 C depending on readout of Tempera...

Page 23: ...ature Reference Source power switch is on Camera and Temperature Reference Source powered on for several minutes Ensure temperatures on back of Temperature Reference Source match Restart Computer Camera is away from direct sunlight or reflective light Plain background Try newer version of IR Flash STM software ...

Page 24: ...ning Institute In addition to providing custom germanium silica and sapphire optics we also build windows for enclosures as well as custom pan and tilt units We can even provide customizable explosion proof systems Our knowledge and experience stems from years of using infrared imaging and temperature measurement instruments to provide solutions to managers engineers scientists inspectors and oper...

Reviews: