ICI FM PLUS P Series User Manual Download Page 16

FM PLUS P SERIES IR CAMERA USER MANUAL

16

9. Cleaning and Maintenance

9-1 Cleaning the Germanium Lens

Do not use corrosive chemicals on the optical glass components. The germanium window 

surface is coated with anti-reflection coating. Dust, grease, and fingerprints will produce 

harmful substances and lead to a decline in performance, or cause scratches. If dirt is found, 

please use the following methods:

1. 

Use a blown balloon or a soft brush to clean the lens surface to avoid dust particles 

scratching the anti-reflection film on lens surface during the wiping process.

2.  Use a soft cotton or microfiber cloth or lens wiping paper and dip in distilled water. 

Gently wipe the lens surface from the middle to the edge, paying attention to not crack 

the lens, or use too much liquid. If the lens is still not clean, replace the cloth and repeat 

the wiping process.

9-2 Disinfecting the Camera Surface

Do not use corrosive cleaning solutions on the optical glass components. 

It is recommended 

to disinfect the camera surface regularly with a non-corrosive sanitizing product. Follow the 

directions provided by the manufacturer of the cleaning solution. Adhere to the sanitation 

protocols and cleaning schedule set forth by the employer. 

9-3 Device Calibration

It is recommended to have device(s) re-calibrated annually. Contact customer service to 

schedule maintenance.

9-4 Storage

When the equipment is not in use, the Temperature Reference Source should be  placed in a 

dust-free and moisture-free environment with a stable temperature and humidity. 

Summary of Contents for FM PLUS P Series

Page 1: ...rt Control THIS DEVICE IS INTENDED FOR ADJUNCTIVE USE WITH OTHER CLINICAL DIAGNOSTIC PROCEDURES TO MEASURE HUMAN BODY TEMPERATURE VIA NON CONTACT SKIN MEASUREMENTS VISUALIZED FROM THE HUMAN FACE NOT M...

Page 2: ...TE BUT ARE PRESENTED WITHOUT WARRANTY OF ANY KIND NOTWITHSTANDING ANY OTHER WARRANTY HEREIN ALL DOCUMENT FILES AND SOFTWARE ARE PROVIDED AS IS WITH ALL FAULTS ICI DISCLAIMS ALL WARRANTIES EXPRESSED OR...

Page 3: ...ty Information 3 Intended Use 4 Technical Specifications 5 Structure 6 Package Contents 7 Installation Instructions 8 Computer Setup 9 Cleaning and Maintenance 10 Site and System Validation Checklist...

Page 4: ...with care during use storage and transportation to prevent dangerous actions such as the device being hit by external force or falling from heights During transportation and storage the original pack...

Page 5: ...nge 20 C to 50 C 68 F to 122 F Operation Range 0 C to 50 C 32 F to 122 F Storage Range 40 C to 60 C 40 F to 140 F Humidity 5 to 95 non condensing Accuracy 0 3 C 0 54 F Pixel Operability 99 Dimensions...

Page 6: ...IR CAMERA USER MANUAL 6 5 2 Definitions of Housing Interface 1 2 1 RS485 output reserved 2 COM port reserved 3 CVBS_OUT output reserved 4 DC 12V power supply 5 RJ 45 Internet access 3 4 5 5 Structure...

Page 7: ...USER MANUAL 7 6 Package Contents ENSURE ALL SYSTEM EQUIPMENT AND COMPONENT ITEMS ARE PRESENT BEFORE BEGINNING INSTALLATION Temperature Reference Ethernet Cable FM Series IR Camera Power Cord x2 Softwa...

Page 8: ...E AIRFLOW FROM DOORWAYS AND AIR CONDITIONING VENTILATION SYSTEMS THE SELECTED AREA SHOULD HAVE A STABLE AMBIENT TEMPERATURE BETWEEN 20 C AND 24 C 68 F TO 75 2 F AND RELATIVE HUMIDITY RANGE FROM 10 TO...

Page 9: ...tripod using the 20 mount MAKE SURE TRIPODS DO NOT BLOCK THE DIRECT PATH OF PERSON S TO BE IMAGED TO ENSURE THE EQUIPMENT WILL NOT BE MOVED OR KNOCKED DOWN USING A DIVIDING BARRIER WILL HELP KEEP TRIP...

Page 10: ...end of the Ethernet cable into the FM Series IR Camera Ethernet port cable Plug the other end of the Ethernet cable into a computer Ethernet port G H Plug power cord into the power adapter Plug the p...

Page 11: ...W AT LEAST 45 MINUTES FOR THE TEMPERATURE REFERENCE SOURCE TO WARM UP THIS WILL PROVIDE THE MOST ACCURATE SKIN TEMPERATURE MEASUREMENTS SOFTWARE ARRIVES INSTALLED ON COMPUTERS AND TABLETS PURCHASED FR...

Page 12: ...ed with secondary evaluation methods e g an NCIT or clinical grade contact thermometer Signage should instruct for the removal of glasses headwear and masks Read the current government guidance regard...

Page 13: ...Subnet Mask should read 255 255 255 0 If it does not click in the box and fill in the correct number 10 Click Ok 8 1 Computer Setup The FM 320 Plus FM 640 Plus comes configured with static IP addresse...

Page 14: ...IR Flash Skin Temperature Monitor Software IR Flash STM Software arrives installed on computers and tablets purchased from ICI as complete systems and it will launch automatically on startup A copy o...

Page 15: ...net be sure to locate the computer behind a firewall Install security updates to the computer s operating system in accordance with your organization s policies Do not use the system s computer for ot...

Page 16: ...e the lens surface from the middle to the edge paying attention to not crack the lens or use too much liquid If the lens is still not clean replace the cloth and repeat the wiping process 9 2 Disinfec...

Page 17: ...light Warm or cold conductive airflow HVAC vents intakes Any radiant energy from electrical sources Direct or indirect lighting on individuals being screened Room temperature is 20 C 24 C 68 F 75 F Re...

Page 18: ...ment on the floor V X or a set of footprints The distance is unique to each screening system Ensure the line of sight from the temperature reference source to the camera is never blocked by an individ...

Page 19: ...od or other stable stand in front of background and positioned for visibility within the frame of the image Temperature is set at 35 C 95 F or 37 C 98 6 F Warm up time is 15 minutes 10 7 Software Set...

Page 20: ...readings are incorrect or facial recognition is suboptimal Close and reopen software Check that software settings are correct Fixed temperature of 35 C 95 F or 37 C 98 6 F depending on temperature re...

Page 21: ...eral minutes Ensure temperatures on back of temperature reference source match Camera is away from direct sunlight or reflective light Plain background Try IR Flash STM software If already using try a...

Page 22: ...Institute In addition to providing custom germanium silica and sapphire optics we also build windows for enclosures as well as custom pan and tilt units We can even provide customizable explosion pro...

Reviews: