ICI FM PLUS P Series User Manual Download Page 14

FM PLUS P SERIES IR CAMERA USER MANUAL

14

8-2-2 Inspection Quarantine Software

First generation FM Plus Series systems shipped with Inspection Quarantine software 

installed 

on computers and tablets purchased from ICI as complete systems

. ICI recommends 

upgrading to the most current version of IR Flash STM as ICI no longer offers support  for 

Inspection Quarantine Software. Uninstall the software as follows:

1. 

Go to Window Control Panel.

2.  Click Programs.

3.  Click Uninstall a program.

4.  Find the software in the list and click it.

5.  Click Uninstall/Change.

6.  Click Yes to confirm you want to uninstall.

7.  Click Ok.

8-2-1 IR Flash Skin Temperature Monitor Software (IR Flash STM) 

Software arrives installed on computers and tablets purchased from ICI as complete 

systems and it will launch automatically on startup.  A copy of the software is provided on 

the Software USB Drive. 

Documentation can be found online on the IR Flash STM web page 

under the Downloads section or at this address:

https://infraredcameras.com/ir-flash-stm-manual/

• 

Processor: 

i5 or above (Quad Core or better)

• 

RAM: 

4 GB or above

• 

OS:

 Windows 8/8.1/10

• 

Hard Drive: 

256 GB or above

• 

Resolution:

 1920 x 1080

8-2 Software

ICI RECOMMENDS UPGRADING TO THE MOST CURRENT VERSION 

OF IR FLASH STM AS SUPPORT FOR INSPECTION QUARANTINE 

SOFTWARE IS NO LONGER OFFERED. UNINSTALL THE SOFTWARE 

USING THE ABOVE INSTRUCTIONS. CONTACT  CUSTOMER 

SERVICE FOR THE MOST UP-TO-DATE VERSION OF IR FLASH STM.

Summary of Contents for FM PLUS P Series

Page 1: ...rt Control THIS DEVICE IS INTENDED FOR ADJUNCTIVE USE WITH OTHER CLINICAL DIAGNOSTIC PROCEDURES TO MEASURE HUMAN BODY TEMPERATURE VIA NON CONTACT SKIN MEASUREMENTS VISUALIZED FROM THE HUMAN FACE NOT M...

Page 2: ...TE BUT ARE PRESENTED WITHOUT WARRANTY OF ANY KIND NOTWITHSTANDING ANY OTHER WARRANTY HEREIN ALL DOCUMENT FILES AND SOFTWARE ARE PROVIDED AS IS WITH ALL FAULTS ICI DISCLAIMS ALL WARRANTIES EXPRESSED OR...

Page 3: ...ty Information 3 Intended Use 4 Technical Specifications 5 Structure 6 Package Contents 7 Installation Instructions 8 Computer Setup 9 Cleaning and Maintenance 10 Site and System Validation Checklist...

Page 4: ...with care during use storage and transportation to prevent dangerous actions such as the device being hit by external force or falling from heights During transportation and storage the original pack...

Page 5: ...nge 20 C to 50 C 68 F to 122 F Operation Range 0 C to 50 C 32 F to 122 F Storage Range 40 C to 60 C 40 F to 140 F Humidity 5 to 95 non condensing Accuracy 0 3 C 0 54 F Pixel Operability 99 Dimensions...

Page 6: ...IR CAMERA USER MANUAL 6 5 2 Definitions of Housing Interface 1 2 1 RS485 output reserved 2 COM port reserved 3 CVBS_OUT output reserved 4 DC 12V power supply 5 RJ 45 Internet access 3 4 5 5 Structure...

Page 7: ...USER MANUAL 7 6 Package Contents ENSURE ALL SYSTEM EQUIPMENT AND COMPONENT ITEMS ARE PRESENT BEFORE BEGINNING INSTALLATION Temperature Reference Ethernet Cable FM Series IR Camera Power Cord x2 Softwa...

Page 8: ...E AIRFLOW FROM DOORWAYS AND AIR CONDITIONING VENTILATION SYSTEMS THE SELECTED AREA SHOULD HAVE A STABLE AMBIENT TEMPERATURE BETWEEN 20 C AND 24 C 68 F TO 75 2 F AND RELATIVE HUMIDITY RANGE FROM 10 TO...

Page 9: ...tripod using the 20 mount MAKE SURE TRIPODS DO NOT BLOCK THE DIRECT PATH OF PERSON S TO BE IMAGED TO ENSURE THE EQUIPMENT WILL NOT BE MOVED OR KNOCKED DOWN USING A DIVIDING BARRIER WILL HELP KEEP TRIP...

Page 10: ...end of the Ethernet cable into the FM Series IR Camera Ethernet port cable Plug the other end of the Ethernet cable into a computer Ethernet port G H Plug power cord into the power adapter Plug the p...

Page 11: ...W AT LEAST 45 MINUTES FOR THE TEMPERATURE REFERENCE SOURCE TO WARM UP THIS WILL PROVIDE THE MOST ACCURATE SKIN TEMPERATURE MEASUREMENTS SOFTWARE ARRIVES INSTALLED ON COMPUTERS AND TABLETS PURCHASED FR...

Page 12: ...ed with secondary evaluation methods e g an NCIT or clinical grade contact thermometer Signage should instruct for the removal of glasses headwear and masks Read the current government guidance regard...

Page 13: ...Subnet Mask should read 255 255 255 0 If it does not click in the box and fill in the correct number 10 Click Ok 8 1 Computer Setup The FM 320 Plus FM 640 Plus comes configured with static IP addresse...

Page 14: ...IR Flash Skin Temperature Monitor Software IR Flash STM Software arrives installed on computers and tablets purchased from ICI as complete systems and it will launch automatically on startup A copy o...

Page 15: ...net be sure to locate the computer behind a firewall Install security updates to the computer s operating system in accordance with your organization s policies Do not use the system s computer for ot...

Page 16: ...e the lens surface from the middle to the edge paying attention to not crack the lens or use too much liquid If the lens is still not clean replace the cloth and repeat the wiping process 9 2 Disinfec...

Page 17: ...light Warm or cold conductive airflow HVAC vents intakes Any radiant energy from electrical sources Direct or indirect lighting on individuals being screened Room temperature is 20 C 24 C 68 F 75 F Re...

Page 18: ...ment on the floor V X or a set of footprints The distance is unique to each screening system Ensure the line of sight from the temperature reference source to the camera is never blocked by an individ...

Page 19: ...od or other stable stand in front of background and positioned for visibility within the frame of the image Temperature is set at 35 C 95 F or 37 C 98 6 F Warm up time is 15 minutes 10 7 Software Set...

Page 20: ...readings are incorrect or facial recognition is suboptimal Close and reopen software Check that software settings are correct Fixed temperature of 35 C 95 F or 37 C 98 6 F depending on temperature re...

Page 21: ...eral minutes Ensure temperatures on back of temperature reference source match Camera is away from direct sunlight or reflective light Plain background Try IR Flash STM software If already using try a...

Page 22: ...Institute In addition to providing custom germanium silica and sapphire optics we also build windows for enclosures as well as custom pan and tilt units We can even provide customizable explosion pro...

Reviews: